Analog Input Specifications |
Measurement functionality |
4-channel voltage measurement, switchable between instantaneous value (waveform) and RMS value |
Input jacks |
Isolated BNC jacks (input resistance: 1 MO; input capacitance: 7 pF) |
Max. rated input-to-ground voltage |
600 V AC/DC, measurement category III
300 V AC/DC, measurement category IV (Inputs are isolated from other instrument circuitry. These figures represent the maximum voltage that can be applied between input channels and the instrument’s enclosure and between input channels themselves without damaging the instrument.) |
Measurement ranges |
10 mV to 100 V/div, 13 ranges, full scale of 10 div
AC voltage that can be measured/displayed using high-speed function: 600 Vrms
Low-pass filter: 5/50/500/5k/50k Hz |
Max. sampling speed |
1 MS/s (4 channels sampled simultaneously) |
Frequency characteristics |
DC to 100 kHz ±3 dB |
Max. rated voltage between input terminals |
600 V AC/DC (Representing the maximum voltage that can be applied between input jacks without damaging the instrument.) |
Analog Input Specifications |
Measurement functionality |
High-speed function (high-speed recording)
Real-time function (real-time recording) |
Max. sampling speed |
1 MS/sec (1 µs period; all channels sampled simultaneously) |
Internal memory capacity |
14 bits x 1m word per channel (1 word = 2 bytes, not expandable) |
External memory |
CF card slot x 1 (up to 2 GB; supported formats: FAT/FAT32)
USB memory x 1 (USB 2.0 Standard-A receptacle) |
Communications interfaces |
USB 2.0 Mini-B receptacle x 1
Functionality: Device configuration and measurement using communications commands; transfer of files from CF/USB memory to a computer (USB drive mode) |
Operating environment (non-condensing) |
Operating temperature range: -10°C to 50°C
Operating humidity range: 80% RH or less at -10°C to 40°C, 60% RH or less at 40°C to 45°C, 50% RH or less at 45°C to 50°C
When powering the instrument with the Z1000 battery: 0°C to 40°C, 80% RH or less
When charging the Z1000 battery: 10°C to 40°C, 80% RH or less |
Storage environment (non-condensing) |
Storage temperature range: -20°C to 60°C
Storage humidity range: 80% RH or less at -20°C to 40°C, 60% RH or less at 40°C to 45°C, 50% RH or less at 45°C to 60°C
Z1000 battery: -20°C to 40°C, 80% RH or less |
Standard compliance |
Safety: EN61010
EMC: EN61326, EN61000-3-2, EN61000-3-3
Vibration resistance: JIS D 1601, Type 1: passenger vehicles; conditions: equivalent to Type A |
Charging function (reference value at 23°C) |
Charge time: Approx. 3 hours (The charging function is activated by connecting the AC adapter while the BATTERY PACK Z1000 is attached to the instrument.) |