(Accuracy guaranteed for one year at 23°C±5°C(73°F±9°F), 80%RH or less. Endurance number of the core opening and closing part: 10000 times) |
Rated primary current |
100 A AC |
Output voltage |
1 mV AC/A |
Amplitude accuracy |
±0.3% rdg.±0.02%f.s. (f.s.:100 A (9695-03),45Hz to 66Hz, at core center) |
Phase accuracy |
Within ±1° (45 Hz to 5 kHz) |
Amplitude frequency characteristics |
Within ±1% at 40 Hz to 5 kHz (deviation from accuracy) |
Effect of conductor position |
Within ±0.5% (deviation from center) |
Effect of external electromagnetic field |
0.1 A equivalent or lower (in an AC electromagnetic field of 400 A/m) |
Maximum permissible input |
130 A continuous (9695-03)(at 45 to 66 Hz, ambient temperature 50°C) |
Temperature coefficient |
0.02%rdg./°C |
Dielectric strength |
3536 V ACrms for 15 sec (between electric circuit and core) |
Maximum rated voltage to earth |
300 V ACrms or lower (Insulated conductor) |
Operating Temperature & Humidity |
0 to 50°C (32 to 122°F), 80%RH or lower (non-condensating) |
Storage Temperature & Humidity |
-10 to 60°C(14 to 140°F), 80%RH or lower (non-condensating) |
Operating Environment |
Indoors, up to 2000 m (6562-ft.) ASL |
Standards applying |
Safety
EN61010-2-032:2002 Type B current sensor
Measurement Category III, Pollution Degree 2
(Anticipated Transient Overvoltage: 4000 V)
EMC
EN61326:1997+A1:1998+A2:2001+A3:2003 |
Measurable conductor diameter |
15 mm (0.59”) or less |
Output terminals |
M3 terminal (Maximum outside diameter: 6.5 mm/0.26”, inside diameter: 3.2 mm/0.13”) |
Cable length |
3 m or less |
Dimensions |
Approx. 50.5W x 58.0H x 18.7D mm (1.99”W x 2.28”H x 0.74”D) (excluding protrusions) |
Mass |
Approx. 50 g (1.8 oz.) |
Accessory |
Instruction Manual |
Option |
9219 CONNECTION CABLE, 9238 CLAMP SENSOR CABLE |
f.s.: maximum display value or scale length
rdg.: reading value (The value currently being measured and indicated on the measuring product) |