Temperature Resistive Probes: |
Pt 50 to 1,000; Ni 100 to 1,000; Cu 10 to 50 |
Example: |
Pt 100: |
Accuracy/one year Measurement |
0.012%R+0.05°C, 0.01°C Resolution |
Accuracy/one year Emission |
0.014%R+0.12°C, 0.02°C Resolution |
Thermocouples: |
13 different types, K, T, J, E, R, S, B, U, L, C, N, Platine, Mo |
Example: |
Type K |
Accuracy/one year Measurement & Emission |
0/1372C = 0.013%R+0.08°C, 0.05°C Resolution, |
DC Current: |
Example: |
up to 50mA with 1µA Resolution |
Accuracy/one year measurement & emission |
0.018%R+2µA |
DC Voltage: |
Four ranges, +/-100mV, 1V, 10V, 50V |
Example: |
+/-100mV with 1µV Resolution |
Accuracy/one year measurement & emission |
0.013%R+3µV (+2µV) |
Resistance: |
Two ranges, 400ohms, 4000ohms |
Example: |
400ohms with 1m ohms resolution |
Accuracy/one year |
0.012%R+10m ohms Measurement |
Frequency and Counting: |
Example: |
20KHz with 0.01Hz |
Accuracy/one year |
0.005%R |
NIST Certification Included |