| Model Number |
TBS1022 |
TBS1042 |
TBS1062 |
TBS1064 |
TBS1102 |
TBS1104 |
TBS1152 |
TBS1154 |
| Bandwidth 1 |
25 MHz |
40 MHz |
60 MHz |
60 MHz |
100 MHz |
100 MHz |
150 MHz |
150 MHz |
| Channels |
2 |
2 |
2 |
4 |
2 |
4 |
2 |
4 |
| Sample rate on each channel |
500 MS/s |
500 MS/s |
1.0 GS/s |
1.0 GS/s |
1.0 GS/s |
1.0 GS/s |
1.0 GS/s |
1.0 GS/s |
| Record length |
2.5k points at all time bases |
| Vertical system — Analog channels |
| Vertical resolution |
8 bits |
| Input sensitivity range |
2 mV to 5 V/div on all models with calibrated fine adjustment |
| DC gain accuracy |
±3%, from 10 mV/div to 5 V/div |
| Maximum input voltage |
300 V RMS CAT II; derated at 20 dB/decade above 100 kHz to 13 V p-p AC at 3 MHz |
| Offset range |
2 mV to 200 mV/div: ±1.8 V
>200 mV to 5 V/div: ±45 V |
| Bandwidth limit |
20 MHz |
| Input coupling |
AC, DC, GND |
| Input impedance |
1 MΩ in parallel with 20 pF |
| Vertical zoom |
Vertically expand or compress a live or stopped waveform |
| Horizontal system — Analog channels |
| Time base range |
5 ns to 50 s/div |
| Time base accuracy |
50 ppm |
| Horizontal zoom |
Horizontally expand or compress a live or stopped waveform |
| Input/Output ports |
| USB interface |
USB host port on front panel supports USB flash drives USB device port on back of instrument supports connection to PC and all
PictBridge®-compatible printers |
| GPIB interface |
Optional |
| Data storage |
| Nonvolatile storage |
| Reference waveform display |
2.5K point reference waveforms |
Waveform storage without
USB flash drive |
2.5K point |
| Maximum USB flash drive size |
64 GB |
Waveform storage with USB
flash drive |
96 or more reference waveforms per 8 MB |
Setups without USB flash
drive |
10 front-panel setups |
| Setups with USB flash drive |
4000 or more front-panel setups per 8 MB |
Screen images with USB flash
drive |
128 or more screen images per 8 MB (the number of images depends on file format selected) |
| Save All with USB flash drive |
12 or more Save All operations per 8 MB
A single Save All operation creates 3 to 9 files (setup, image, plus one file for each displayed waveform)
|
| Acquisition system |
| Acquisition modes |
| Peak Detect |
High-frequency and random glitch capture. Captures glitches as narrow as 12 ns (typical) at all time base settings from 5 μs/div to
50 s/div |
| Sample |
Sample data only |
| Average |
Waveform averaged, selectable: 4, 16, 64, 128 |
| Single Sequence |
Use the Single Sequence button to capture a single triggered acquisition sequence |
| Roll |
At acquisition time base settings of >100 ms/div |
| Trigger system |
| External trigger input |
Included on all models |
| Trigger modes |
Auto, Normal, Single Sequence |
| Trigger types |
|
| Edge (Rising/Falling) |
Conventional level-driven trigger. Positive or negative slope on any channel. Coupling selections: AC, DC, Noise Reject, HF
Reject, LF Reject |
| Video |
Trigger on all lines or individual lines, odd/even or all fields from composite video, or broadcast standards (NTSC, PAL, SECAM) |
| Pulse Width (or Glitch) |
Trigger on a pulse width less than, greater than, equal to, or not equal to, a selectable time limit ranging from 33 ns to 10 s |
| Trigger source |
Two channel models: CH1, CH2, Ext, Ext/5, AC Line
Four channel models: CH1, CH2, CH3, CH4, Ext, Ext/5, AC Line |
| Trigger view |
Displays trigger signal while Trigger View button is depressed. |
| Trigger signal frequency readout |
Provides a frequency readout of the trigger source. |
| Waveform measurements |
| Cursors |
| Types |
Amplitude, Time |
| Measurements |
ΔT, 1/ΔT, ΔV |
| Automatic measurements |
Period, Frequency, +Width, –Width, Rise Time, Fall Time, Max, Min, Peak-to-Peak, Mean, RMS, Cycle RMS, Cursor RMS, Duty
Cycle, Phase, and Delay |
| Waveform math |
| Arithmetic |
Add, Subtract, Multiply |
| Math functions |
FFT |
| FFT |
Windows: Hanning, Flat Top, Rectangular 2048 sample points |
| Sources |
Two channel models: CH1 – CH2, CH2 – CH1, CH1 + CH2, CH1 × CH2
Four channel models: CH1 – CH2, CH2 – CH1, CH1 + CH2, CH1 × CH2, CH3 – CH4, CH4 – CH3, CH3 + CH4, CH3 × CH4 |
| Autoset |
| Autoset menu |
Single-button, automatic setup of all channels for vertical, horizontal, and trigger systems, with undo Autoset |
| Square wave |
Single Cycle, Multicycle, Rising or Falling Edge |
| Sine wave |
Single Cycle, Multicycle, FFT Spectrum |
| Video (NTSC, PAL, SECAM) |
Field: All, Odd, or Even Line: All or Selectable Line Number |
| Autorange |
| |
Automatically adjust vertical and/or horizontal oscilloscope settings when probe is moved from point to point, or when the signal
exhibits large changes. |
| Display system |
| Interpolation |
Sin (x)/x |
| Waveform styles |
Dots, vectors |
| Persistence |
Off, 1 s, 2 s, 5 s, infinite |
| Format |
YT and XY |
| Physical characteristics |
| Dimensions |
mm |
in. |
| |
|
158.0 |
6.22 |
| |
|
326.3 |
12.85 |
| |
|
124.2 |
4.89 |
| Shipping dimensions |
mm |
in. |
| |
|
266.7 |
10.5 |
| |
|
476.2 |
18.75 |
| |
|
228.6 |
9.0 |
| Weight |
kg |
lb. |
| |
|
2.0 |
4.4 |
| |
|
2.2 |
4.9 |
| RM2000B rackmount |
mm |
in |
| |
|
482.6 |
19.0 |
| |
|
177.8 |
7.0 |
| |
|
108.0 |
4.25 |
| Environmental |
|
|
|
| Temperature |
|
| Operating |
0 to +50 ºC |
| Non operating |
–40 to +71 ºC |
| Humidity |
|
|
|
| Operating and non operating |
Up to 85% RH at or below +40 ºC
Up to 45% RH up to +50 ºC
|
| |
| Altitude |
| Operating and non operating |
Up to 3,000 m (9,843 ft.) |
| Regulatory |
|
|
|
| Electromagnetic compatibility |
Meets Directive 2004/108/EC, EN 61326-2-1 Class A; Australian EMC Framework |
| Safety |