FlexChannel inputs |
4 |
Maximum analog channels |
4 |
Maximum digital channels (with optional logic probes) |
32 |
Bandwidth (calculated rise time) |
1 GHz (400 ps) |
DC Gain Accuracy |
< 2 GHz models:
50 Ω: ±1.0%, (±2.0% at ≤ 1 mV/div)
±0.5% of full scale, (±1.0% of full scale at 1 mV/Div and 500 μV/Div Settings)
1 MΩ: ±1.0%, (±2.0% at ≤ 1 mV/div)
±0.5% of full scale, (±1.0% of full scale at 1 mV/Div and 500 μV/Div Settings)
2 GHz models:
50 Ω: ±1.2%, (±2.0% at ≤ 1 mV/div)
±0.6% of full scale, (±1.0% of full scale at 1 mV/Div and 500 μV/Div Settings)
1 MΩ: ±1.0%, (±2.0% at ≤ 1 mV/div)
±0.5% of full scale, (±1.0% of full scale at 1 mV/Div and 500 μV/Div Settings) |
ADC Resolution |
12 bits |
Vertical Resolution |
8 bits @ 6.25 GS/s
12 bits @ 3.125 GS/s
13 bits @ 1.25 GS/s (High Res)
14 bits @ 625 MS/s (High Res)
15 bits @ 312.5 MS/s (High Res)
16 bits @ ≤125 MS/s (High Res) |
Sample Rate |
6.25 GS/s on all analog / digital channels (160 ps resolution) |
Record Length (std.) |
62.5 Mpoints on all analog / digital channels |
Record Length (opt.) |
125, 250, or 500 Mpoints on all analog / digital channels |
Waveform Capture Rate |
>500,000 wfms/s |
Arbitrary/Function Generator (opt.) |
13 predefined waveform types with up to 50 MHz output |
DVM |
4-digit DVM (free with product registration) |
Trigger Frequency Counter |
8-digit frequency counter (free with product registration) |
Vertical system - analog channels |
Bandwidth selections |
50 Ω: 20 MHz, 250 MHz, and the full bandwidth value of your model
1 MΩ: 20 MHz, 250 MHz, 500 MHz |
Input coupling |
DC, AC |
Input impedance |
50 Ω ± 1%
1 MΩ ± 1% with 13.0 pF ± 1.5 pF (< 2 GHz models)
1 MΩ ± 1% with 14.5 pF ± 1.5 pF (2 GHz models) |
Input sensitivity range |
1 MΩ |
500 µV/div to 10 V/div in a 1-2-5 sequence |
50 Ω |
500 µV/div to 1 V/div in a 1-2-5 sequence
Note: 500 μV/div is a 2X digital zoom of 1 mV/div |
Maximum input voltage |
50 Ω: 5 VRMS, with peaks ≤ ±20 V (DF ≤ 6.25%)
1 MΩ: 300 VRMS , CAT II
For 1 MΩ, derate at 20 dB/decade from 4.5 MHz to 45 MHz;
Derate at 14 dB/decade from 45 MHz to 450 MHz; > 450 MHz, 5.5 VRMS |
Vertical system - digital channels |
Number of channels |
8 digital inputs (D7-D0) per installed TLP058 (traded off for one analog channel) |
Vertical resolution |
1 bit |
Maximum input toggle rate |
500 MHz |
Minimum detectable pulse width, typical |
300 ps |
Thresholds |
One threshold per digital channel |
Threshold range |
±40 V |
Threshold resolution |
10 mV |
Threshold accuracy |
± [100 mV + 3% of threshold setting after calibration] |
Input hysteresis, typical |
100 mV at the probe tip |
Input dynamic range, typical |
30 Vpp for Fin ≤ 200 MHz, 10 Vpp for Fin > 200 MHz |
Absolute maximum input voltage, typical |
±42 V peak |
Minimum voltage swing, typical |
400 mV peak-to-peak |
Input impedance, typical |
100 kΩ |
Probe loading, typical |
2 pF |
Horizontal system |
Time base range |
200 ps/div to 1,000 s/div |
Sample rate range |
1.5625 S/s to 6.25 GS/s (real time)
12.5 GS/s to 500 GS/s (interpolated) |
Record length range |
Standard |
1 kpoints to 62.5 Mpoints in single sample increments |
Optional 5-RL-125M |
125 Mpoints |
Optional 5-RL-250M |
250 Mpoints |
Optional 5-RL-500M |
500 Mpoints |
Maximum duration at highest sample rate |
10 ms (std.) or 80 ms (opt.) |
Time base delay time range |
-10 divisions to 5,000 s |
Deskew range |
-125 ns to +125 ns with a resolution of 40 ps |
Timebase accuracy |
±2.5 x 10-6 over any ≥1 ms time interval |
Acquisition system |
Sample |
Acquires sampled values |
Peak Detect |
Captures glitches as narrow as 640 ps at all sweep speeds |
Averaging |
From 2 to 10,240 waveforms |
Envelope |
Min-max envelope reflecting Peak Detect data over multiple acquisitions |
High Res |
Applies a unique Finite Impulse Response (FIR) filter for each sample rate that maintains the maximum bandwidth possible for that sample rate while preventing aliasing and removing noise from the oscilloscope amplifiers and ADC above the usable bandwidth for the selected sample rate.
High Res mode always provides at least 12 bits of vertical resolution and extends all the way to 16 bits of vertical resolution at ≤ 125 MS/s sample rates. |
FastAcq® |
FastAcq optimizes the instrument for analysis of dynamic signals and capture of infrequent events by capturing >500,000 wfms/s (one channel active; >100K wfms/s with all channels active). |
Roll mode |
Scrolls sequential waveform points across the display in a right-to-left rolling motion, at timebase speeds of 40 ms/div and slower, when in Auto trigger mode. |
FastFrame™ |
Acquisition memory divided into segments.
Maximum trigger rate >5,000,000 waveforms per second
Minimum frame size = 50 points
Maximum Number of Frames: For frame size ≥ 1,000 points, maximum number of frames = record length / frame size.
For 50 point frames, maximum number of frames = 1,000,000 |
Waveform measurements |
Cursor types |
Waveform, V Bars, H Bars, V&H Bars, and Polar (XY/XYZ plots only) |
Automatic measurements |
36, of which an unlimited number can be displayed as either individual measurement badges or collectively in a measurement results table |
Amplitude measurements |
Amplitude, Maximum, Minimum, Peak-to-Peak, Positive Overshoot, Negative Overshoot, Mean, RMS, AC RMS, Top, Base, and Area |
Timing measurements |
Period, Frequency, Unit Interval, Data Rate, Positive Pulse Width, Negative Pulse Width, Skew, Delay, Rise Time, Fall Time, Phase, Rising Slew Rate, Falling Slew Rate, Burst Width, Positive Duty Cycle, Negative Duty Cycle, Time Outside Level, Setup Time, Hold Time, Duration N-Periods, High Time, Low Time, Time to Minimum, and Time to Maximum |
Jitter measurements (standard) |
TIE and Phase Noise |
Measurement statistics |
Mean, Standard Deviation, Maximum, Minimum, and Population. Statistics are available on both the current acquisition and all acquisitions |
Reference levels |
User-definable reference levels for automatic measurements can be specified in either percent or units. Reference levels can be set to global for all measurements, per source channel or signal, or unique for each measurement |
Gating |
Screen, Cursors, Logic, Search, or Time. Specifies the region of an acquisition in which to take measurements. Gating can be set to Global (affects all measurements set to Global) or Local (all measurements can have a unique Time gate setting; only one Local gate is available for Screen, Cursors, Logic, and Search actions). |
Measurement plots |
Histogram, Time Trend, Spectrum, Eye Diagram (TIE measurement only), Phase Noise (Phase Noise measurement only) |
Physical characteristics |
Dimensions |
Height: 12.2 in (309 mm), feet folded in, handle to back
Height: 14.6 in (371 mm) feet folded in, handle up
Width: 17.9 in (454 mm) from handle hub to handle hub
Depth: 8.0 in (205 mm) from back of feet to front of knobs, handle up
Depth: 11.7 in (297.2 mm) feet folded in, handle to the back |
Weight |
< 25 lbs (11.4 kg) |
Cooling |
The clearance requirement for adequate cooling is 2.0 in (50.8 mm) on the right side of the instrument (when viewed from the front) and on the rear of the instrument |
Rackmount configuration |
7U (with optional RM5 Rackmount Kit) |
Environmental specifications |
Temperature |
Operating |
+0 °C to +50 °C (32 °F to 122 °F) |
Non-operating |
20 °C to +60 °C (-4 °F to 140 °F) |
Humidity |
Operating |
5% to 90% relative humidity (% RH) at up to +40 °C
5% to 55% RH above +40 °C up to +50 °C, noncondensing, and as limited by a maximum wet-bulb temperature of +39 °C |
Non-operating |
5% to 90% relative humidity (% RH) at up to +40 °C
5% to 39% RH above +40 °C up to +50 °C, noncondensing, and as limited by a maximum wet-bulb temperature of +39 °C |
Altitude |
Operating |
Up to 3,000 meters (9,843 feet) |
Non-operating |
Up to 12,000 meters (39,370 feet) |
EMC, Environmental, and Safety |
Regulatory |
CE marked for the European Union and UL approved for the USA and Canada RoHS compliant |