MSO44 |
MSO46 |
FlexChannel inputs |
4 |
6 |
Maximum analog channels |
4 |
6 |
Maximum digital channels (with optional logic probes) |
32 |
48 |
Auxiliary Trigger Input |
≤300 V RMS (Edge Trigger only) |
Bandwidth (calculated rise time) |
200 MHz, 350 MHz, 500 MHz, 1 GHz, 1.5 GHz |
DC Gain Accuracy |
50 Ω: ±1%, (±2.5% at 1 mV/Div and 500 µV/Div settings), de-rated at 0.100%/°C above 30 °C 1 MΩ and 250 kΩ: ±1.0%, (±2.0% at 1 mV/Div and 500 µV/Div settings) |
ADC Resolution |
12 bits |
Vertical Resolution |
8 bits @ 6.25 GS/s
12 bits @ 3.125 GS/s
13 bits @ 1.25 GS/s (High Res)
14 bits @ 625 MS/s (High Res)
15 bits @ 312.5 MS/s (High Res)
16 bits @ ≤125 MS/s (High Res) |
Sample Rate |
6.25 GS/s on all analog / digital channels (160 ps resolution) |
Record Length (std.) |
31.25 Mpoints on all analog / digital channels |
Record Length (opt.) |
62.5 Mpoints on all analog / digital channels |
Waveform Capture Rate, typical |
>500,000 wfms/s |
Arbitrary/Function Generator (opt.) |
13 predefined waveform types with up to 50 MHz output |
DVM |
4-digit DVM (free with product registration) |
Trigger Frequency Counter |
8-digit frequency counter (free with product registration) |
Bandwidth selections |
50 Ω: 20 MHz, 250 MHz, and the full bandwidth value of your model
1 MΩ: 20 MHz, 250 MHz, 500 MHz |
Input coupling |
DC, AC |
Input impedance |
50 Ω ± 1%
1 MΩ ± 1% with 13.0 pF ± 1.5 pF |
1 MΩ |
500 µV/div to 10 V/div in a 1-2-5 sequence |
50 Ω |
500 µV/div to 1 V/div in a 1-2-5 sequence
Note: 500 μV/div is a 2X digital zoom of 1 mV/div or a 4x digital zoom of 2 mV/div, depending on the instrument bandwidth
configuration |
Maximum input voltage |
50 Ω: 5 VRMS, with peaks ≤ ±20 V (DF ≤ 6.25%)
1 MΩ: 300 VRMS
For 1 MΩ, derate at 20 dB/decade from 4.5 MHz to 45 MHz;
Derate at 14 dB/decade from 45 MHz to 450 MHz; > 450 MHz, 5.5 VRMS |
Number of channels |
8 digital inputs (D7-D0) per installed TLP058 (traded off for one analog channel) |
Vertical resolution |
1 bit |
Minimum detectable pulse width, typical |
1 ns |
Thresholds |
One threshold per digital channel |
Threshold range |
±40 V |
Threshold resolution |
10 mV |
Threshold accuracy |
± [100 mV + 3% of threshold setting after calibration] |
Input hysteresis, typical |
100 mV at the probe tip |
Input dynamic range, typical |
30 Vpp for Fin ≤ 200 MHz, 10 Vpp for Fin > 200 MHz |
Absolute maximum input voltage, typical |
±42 V peak |
Minimum voltage swing, typical |
400 mV peak-to-peak |
Input impedance, typical |
100 kΩ |
Probe loading, typical |
2 pF |
Time base range |
200 ps/div to 1,000 s/div |
Sample rate range |
1.5625 S/s to 6.25 GS/s (real time)
12.5 GS/s to 500 GS/s (interpolated) |
Standard |
1 kpoints to 31.25 Mpoints in single sample increments |
Option 4-RL-1 |
62.5 Mpoints |
Maximum duration at highest sample rate |
5 ms (std.) or 10 ms (opt.) |
Time base delay time range |
-10 divisions to 5,000 s |
Deskew range |
-125 ns to +125 ns with a resolution of 40 ps |