High Resolution
For frequency, period and frequency ratio functions the instrument uses a reciprocal counting technique to provide high resolution at all frequencies. Eight significant digits of answer are produced in a 1 second measurement time, nine digits in 10s and ten digits in 100s with a granularity of less than 2 counts in the least significant digit.
Flexible Signal Conditioning
Input A has configurable coupling (AC or DC), input impedance (1MW or 50W), attenuation (1:1 or 5:1), threshold (fully variable) and active edge, and can be used for frequencies in the range 0.001Hz to >125MHz. Input B is a nominal 50W input for frequencies in the range 80MHz to > 3GHz.
Input C (TF960 only) uses a standard N connector and has a nominal 50W input for frequencies in the range 1.8GHz to > 6GHz.
Multiple Measurement Functions
The TF930 and TF960 can measure frequency, period, pulse width, duty cycle and frequency ratio, as well as event counting (totalise).
Remote Control and Read-Back via USB
The TF930 and TF960 incorporate a USB interface which allows it to be remotely controlled using RS232 protocol via a computer’s USB port. The remote commands of their predecessor instrument, the TF830, are compatible with the command set.
Battery Operation when Needed
The TF930 and TF960 operate from internal rechargeable NiMH batteries which give typically 24 hours operating life. The universal AC charger supplied will recharge the batteries in less than 4 hours and can be used for continuous AC operation. The instruments can also be powered from a standard USB port.