Insulation ranges |
Nominal Test Voltage |
1000 V, 500 V, 250 V (d.c.) |
Measuring Range |
All instruments |
10 k½ - 999 M½ on all ranges |
Short Circuit Current |
1.5 mA nominal |
Test Current on Load |
1 mA at min. pass values of insulation (as specified in BS7671, HD 384 and IEC 364) |
Accuracy (at 20° C) |
±3%, ±2 digits |
Continuity ranges |
Measuring Range |
0,01 ½ - 99,9 ½ (0 -50 ½ on analogue scale) |
Open Circuit Voltage |
5 V ±1 V |
Short Circuit Current |
200 - 215 mA |
Terminal Characteristics |
|
Accuracy (at 20° C) |
±3% ±2 digits |
Zero Offset Adjust |
0 - 9 ½ |
Continuity Buzzer |
Adjustable 1 ½ to 20 ½ Response time <20 ms |
Resistance range MIT320/330 |
(can be used for diode testing) |
Measuring Range |
10 ½ - 1 M½ |
Open Circuit Voltage |
5 V |
Short Circuit Current |
1.5 mA |
Accuracy (at 20° C) |
±5% ±2 digits up to 100 k½ |
Voltage range |
Measuring Range |
0 - 600 V a.c. (50/60 Hz) or d.c. |
Accuracy (at 20° C) |
d.c. or a.c. (50/60 Hz) ±1%, ±2 digits |
Temperature coefficient |
<0,1% per °C on all ranges. |
Default voltmeter |
(all modes except off) |
>25 V a.c. or d.c. is applied display will operate as a voltmeter. |
Test inhibit |
If more than 50 volts is detected, testing will be inhibited. |
AutoPower Down |
Auto power down operates after 10 minutes if left in standby mode. |
Temperature and humidity |
Operating Range |
-10°C to +60°C |
Operating Humidity |
93% R.H. at +40°C max.: |
Storage Range |
-25°C to +70°C |
Test Result Storage (1000 results) |
Job Number |
255 locations |
Distribution Board no. |
99 locations |
Circuits |
99 locations |
Phases |
P1, P2, P3 |
Last recorded result may be recalled to the display.
All data can be downloaded to a PC using Download Manager (supplied) or Megger Powersuite Professional of Windows software |
Communications |
USB interface |
Environmental Protection |
IP54 |
Fuses |
Terminals 500 mA (F) 600 V, 32 x 6 mm Ceramic HBC 10 kA minimum. |
Display shows if fuse is ruptured. |
Safety |
Meets the requirements of IEC61010-1 Cat III 600 V phase to earth. |
Automatic discharge |
After an insulation test the item under test will be discharged automatically. Any voltage present will be indicated on the display so that the discharge can be monitored. |
Power supply |
Battery |
8 x 1,5 V cells IEC LR6 type(AA alkaline).
Rechargeable NiCd or NiMH cells may be used.
Battery condition is constantly shown on the display as a four-section bargraph. |
Battery Life |
5000 consecutive tests (5 seconds per test) on any test using 2Ah batteries. Weight |
All units |
980 gms |
Dimensions |
All units: 203 x 148 x 78 mm (8 x 5.7 x 3.2 inches) |
E.M.C. |
In accordance with IEC61326 including amendment No.1 |