ISOLATION (low–earth) |
22VDC max |
PROGRAMMING |
IEEE-488.2 (SCPI) |
USER-DEFINABLE POWER-UP STATES |
5 (4 for Models 2302-PJ and 2306-PJ). |
REAR PANEL CONNECTORS |
Two (one for Models 2302, 2302-PJ) 8-position quick disconnect terminal block for output (4), sense (2), and DVM (2). |
TEMPERATURE COEFFICIENT
(outside 23°C ±5°C) |
Derate accuracy specification by (0.1 × specification)/°C |
OPERATING TEMPERATURE |
0° to 50°C (Derate to 70%). 0° to 35°C (Full power) |
STORAGE TEMPERATURE |
–20° to 70°C. |
HUMIDITY |
<80% @ 35°C non-condensing. |
DISPLAY TYPE |
2-line × 16 character VFD |
REMOTE DISPLAY/KEYPAD OPTION |
Disables standard front panel |
DIMENSIONS |
89mm high × 213mm wide × 411mm deep (3 1 / 2 in × 8 3 / 8 in × 16 3 / 16 in) |
NET WEIGHT |
3.2kg (7.1 lbs) |
SHIPPING WEIGHT |
5.4kg (12 lbs) |
INPUT POWER |
100–120V AC/220–240V AC, 50 or 60Hz (auto detected at power-up |
POWER CONSUMPTION |
150VA max |
WARRANTY |
Two years parts and labor on materials and workmanship |
EMC |
Conforms with European Union Directive 89/336/EEC, EN 55011, EN 50082-1, EN 61000-3-2 and 61000-3-3, FCC part 15 class B. 2302-PJ, 2306-PJ: Conforms with European Union Directive 89/336/EEC. |
SAFETY |
Conforms with European Union Directive 73/23/EEC, EN 61010-1. |
AC LINE LEAKAGE CURRENT |
450µA @ 110VAC, typ.; 600µA @ 220V, type |
RELAY CONTROL PORT |
4-channel, each capable of 100mA sink, 24V max. Total port sink capacity (all 4 combined) is 250mA max. Accepts DB-9 male plug. |
OUTPUT VOLTAGE |
0 to +15VDC |
OUTPUT ACCURACY |
±(0.05% + 3mV) |
PROGRAMMING RESOLUTION |
1mV |
READBACK ACCURACY |
±(0.05% + 3mV) |
READBACK RESOLUTION |
1mV |
OUTPUT VOLTAGE SETTLING TIME |
5ms to within stated accuracy |
LOAD REGULATION |
0.01% + 2mV |
LINE REGULATION |
0.5mV |
STABILITY |
0.01% + 0.5mV |
MEASUREMENT TIME CHOICES |
0.01 to 10 PLC 7 , in 0.01PLC steps |
AVERAGE READINGS |
1 to 10 |
READING TIME |
31ms, typical |
READBACK ACCURACY |
High Bandwidth |
Low Bandwidth |
Transient Recovery Time |
<40µs or <60µs |
<80µs or <100µs |
Transient Voltage Drop |
<75mV or <100mV |
<250mV or <400mV |
REMOTE SENSE |
1V max. drop in each lead. Add 2mV to the voltage load regulation specification for each 1V change in the negative output lead due to load current change. Remote sense required. Integrity of connection continually monitored. If compromised, output will turn off automatically once settable window (±0 to ±8 volts) around normal voltage exceeded. |
RANGE |
0 to 1.00? in 0.01? steps. Value can be changed with output on |
OUTPUT CURRENT |
0–4V |
5A max |
>4V |
I MAX = 60W/(V OUT +6) (not intended to be operated in parallel) |
SOURCE COMPLIANCE ACCURACY |
±(0.16% + 5mA) |
PROGRAMMED SOURCE COMPLIANCE RESOLUTION |
1.25mA |
READBACK ACCURACY |
5A Range |
±(0.2% + 200µA) |
5mA Range |
±(0.2% + 1µA) |
READBACK RESOLUTION |
5A Range |
100µA |
5mA Range |
0.1µA |
CURRENT SINK CAPACITY |
0–5V |
3A max |
5V–15V |
Derate 0.2A per volt above 5V. Compliance setting controls sinking |
LOAD REGULATION |
0.01% + 1mA |
LINE REGULATION |
0.5mA |
STABILITY |
0.01% + 50µA |
MEASUREMENT TIME CHOICES |
0.01 to 10 PLC , in 0.01PLC steps |
AVERAGE READINGS |
1 to 10 |
READING TIME |
31ms, typical |
TRIGGER LEVEL |
5A Range |
5mA to 5A, in 5mA steps |
1A Range |
1mA to 1A, in 1mA steps |
100mA Range |
0.1mA to 100mA, in 100µA steps |
TRIGGER DELAY |
0 to 100ms, in 10µs steps |
INTERNAL TRIGGER DELAY |
15µs |
HIGH/LOW/AVERAGE MODE |
Measurement Aperture Settings |
33.3µs to 833ms, in 33.3µs steps |
Average Readings |
1 to 100 |
PULSE CURRENT MEASUREMENT ACCURACY 11 (2 Years, 23°C ±5°C) |
APERTURE |
ACCURACY ±(% reading + offset + rms noise 10 ) |
<100 µs |
0.2% + 900 µA + 2 mA |
100 µs – 200 µs |
0.2% + 900 µA + 1.5mA |
200 µs – 500 µs |
0.2% + 900 µA + 1 mA |
500 µs – <1 PLC |
0.2% + 600 µA + 0.8mA |
1 PLC |
0.2% + 400 µA + 0 mA |
>1 PLC |
0.2% + 400 µA + 100 µA |
MEASUREMENT APERTURE |
33.3µs |
CONVERSION RATE |
3650/second, typical |
INTERNAL TRIGGER DELAY |
15µs |
NUMBER OF SAMPLES |
1 to 5000 |
TRANSFER SAMPLES ACROSS IEEE BUS IN BINARY MODE |
4800 bytes/s, typical |
MEASUREMENT TIME |
850ms (840ms) to 60 seconds in 1ms steps |
INPUT VOLTAGE RANGE |
–5 to +30VDC |
INPUT IMPEDANCE |
2M? typical |
MAXIMUM VOLTAGE (either input terminal)
WITH RESPECT TO OUTPUT LOW |
–5V, +30V |
READING ACCURACY |
±(0.05% + 3mV) |
READING RESOLUTION |
1mV |
CONNECTOR |
HI and LO input pair part of Output #1’s terminal block |
MEASUREMENT TIME CHOICES |
0.01 to 10 PLC 7 , in 0.01PLC steps |
AVERAGE READINGS |
1 to 10 |