Max. applied measurement voltage |
±300 V DC rated input voltage
±300 V DC maximum rated voltage above ground |
±60 V DC rated input voltage
±70 V DC maximum rated voltage above ground |
Resistance Measurement ranges |
3 mO (Max. display 3.1000 mO, resolution 0.1 µO) to 3000 O (Max. display 3100.0 O, resolution 100 mO), 7 ranges
Accuracy: 30 mO to 3000O ranges, ±0.5 % rdg. ±5 dgt. (Add ±3 dgt. for EX.FAST, or ±2 dgt. for FAST and MEDIUM)
3 mO range, ±0.5 % rdg. ±10 dgt. (Add ±30 dgt. for EX.FAST, or ±10 dgt. for FAST, or ±5 dgt. for MEDIUM)
Testing source frequency: 1kHz ±0.2Hz, testing current: 100 mA (3 mO range) to 10 µA (3000O range)
Open-circut Voltage: 25 V peak (3/30 mO ranges), 7 V peak (300 mO range), 4 V peak (3O to 3000O ranges) |
Voltage Measurement ranges |
DC 6 V (resolution 10 µV) to DC 300 V (resolution 1 mV), 3 ranges |
DC 6 V (resolution 10 µV) to DC 60 V (resolution 100 µV), 2 ranges |
Voltage Measurement ranges |
Accuracy: ±0.01 % rdg. ±3 dgt. (Add ±3 dgt. for EX.FAST, or ±2 dgt. for FAST and MEDIUM) |
Display |
31000 full digits (resistance), 600000 full digits (voltage), LED |
Sampling rate |
Four steps
4 ms(Extra-FAST), 12 ms(FAST), 35 ms(Medium), 150 ms(Slow)
(Typ., sampling time depends on supply frequency settings and function.) |
Measurement time |
Response time + sampling rate, approx. 10 ms for measurements
(Response time depends on reference values and the measurement object.) |
Comparator functions |
Setting: Upper and lower limit
Output: 3 levels (Hi, In, Lo) or (Pass, Fail), Open-collector(35V, DC50mA max), Display, Dual audible indicator |
Analog Output |
Mesured resistance (displayed value, from 0 to 3.1V DC, -01 suffix models only) |
Interface |
External I/O, RS-232C, Printer (RS-232C), GP-IB(-01 suffix models only) |
Power supply |
100 to 240 V AC, 50/60 Hz, 30 VA max. |
Dimensions, mass |
215 mm (8.46 in) W × 80 mm (3.15 in) H × 295 mm (11.61 in) D, 2.4 kg (84.7 oz) |