Accuracy guaranteed for one year at 23 ± 5°C (73 ± 9°F), 80%RH or less (Opening and Closing of the Sensor: Maximum 10000 times). |
Rated primary current |
10 A AC |
Output voltage |
100 mV AC/A |
Amplitude accuracy |
± 1.0% rdg. ± 0.005% f.s. (f.s.: 10 A at the clamp core center, 45 to 66 Hz) |
Phase accuracy |
within ± 5° (at 50 Hz or 60 Hz) |
Amplitude frequency characteristics |
Within ± 5% at 40 Hz to 5 kHz (deviation from accuracy) |
Maximum input current |
10 A continuous at 45 Hz to 66 Hz (Ambient temperature: 50°C) |
Effect of conductor position |
within ± 0.1% (deviation from center) |
Effect of external magnetic fields |
Max. 7.5 mA (400 A AC/m corresponds) |
Remaining electric current character |
1 mA or less (10 A AC, 50/60 Hz on forward and return lines with 10 mm or less line spacing) |
Dielectric strength |
2210 Vrms for 1 minute (between electric circuit and core, core and case) |
Maximum rated voltage to earth |
300 Vrms or less |
Operating temperature and humidity range |
0 to 50°C (32 to 122°F) 80%RH or less (no condensation) |
Storage temperature and humidity range |
-10 to 60°C (14 to 140°F) 80%RH or less (no condensation) |
Location for use |
Altitude up to 2000 m (6562 feet), Indoors |
Standards applying |
(Safety)
EN61010-2-032:2002
Type B current sensor
Measurement Category III, Pollution
Degree 2 (anticipated transient overvoltage 2210 V)
(EMC)
EN61326:1997+A1:1998+A2:2001
+A3:2003 |
Diameter of measurable conductor |
30 mm (1.18") or less |
Cable length |
Isolated BNC, Approx. 3 m (118.11") |
External dimensions |
Approx. 60W x 112.5H x 23.6D mm (2.36"W x 4.43"H x 0.93"D, excluding protrusions) |
Mass |
Approx. 160 g (5.6 oz.) |
Accessories |
Instruction manual (1) |