Time axis |
100 μs to 5 min/div, 20 ranges, time axis resolution 100 points/div,
time axis zoom: ×2 to ×10 in 3 stages, compression: 1/2 to 1/1,000 in 9 stages, Auto roll mode display at 50 ms/div or slowely range |
Sampling speed |
1/100th of time axis range, (1 μs period maximum, simultaneous sampling in all channels) |
Recording length |
Ten settings from 20 to 20,000 div, or continuous
(limited by timebase, only the last 20,000 div are saved) |
Pre-trigger |
Records waveforms prior to trigger events, from 0 to 100% of the specified recording length |
Screen types |
Split screen (none, only 1 screen), X-Y screen (none, but possible at use with the supplied PC software), Waveform or numerical logging (switched), Voltage axis zoom (×2 to ×10), compression (×1/2 to ×1/5) |
Numerical display |
Instantaneous value or RMS value display (only DC and 50/60 Hz)
Reflesh rate: 0.5 sec, Sampling speed: 10 kS/sec, 4 digits: (the lowest digit displays as 0 for values 0 to 4, and 5 for values 5 to 9)
Voltage axis range 10 mV, 50 mV, 100 mV, 500 mV, 1 V, 5 V, 10 V, 50 V/div, auto ranging,
Accuracy : ±2.5 % of reading ±5 digits |
Numerical calculation |
Up to four simultaneous calculations (common to all channels), calculation results are saved to CF card,
Calculation contents: average, peak, maximum and minimum values, RMS, period and frequency
Calculation range: specified by A/B cursors or whole recording length |
Saving function |
Setting configurations, measurement data (binary and text), screen data (compressed bitmap format), numerical calculation results, thinned data saving (text) |
Cursor readout |
Trace, vertical and horizontal |
Scaling function |
Selectable by model (clamp or Differential Probe 9322), specified conversion ratio (output ratio, division ratio), 2-point setting method |
Comment entry |
Title comments can be entered for each channel (including logic) |
Screen capture |
The displayed screen is saved to CF card as a compressed bitmap |
Gauges |
Vertical axis gauges can be displayed for two channels on the waveform screen |
Preserve starting conditions |
If power fails while measuring, measurement can be automatically resumed when power is restored |
Auto save |
Included |
Scroll bar |
To jump to a specific waveform location |
Waveform monitor |
Setting is possible while monitoring waveforms on the Settings screen |
Logic signal view |
In 4-bit units, four selectable display positions |