Model LR8534 Specifications |
Connectable instrument |
Memory HiLogger LR8450-01 (Wireless LAN model) |
Control communications method |
Connect wirelessly via Wireless LAN Adapter Z3230 (included)
Wireless LAN (IEEE 802.11b/g/n)
Communication range: 30 m (line of sight)
Encryption function: WPA-PSK/WPA2-PSK, TKIP/AES
Available number of channels: 1 to 11 |
Communications buffer memory |
4 M word (volatile memory)
Saves data in the event of a communications error. Data is resent when communications are restored. |
Operating temperature and humidity |
-20°C to 55°C, 80% RH (no condensation)
(Charging temperature range: 5°C to 35°C) |
Storage temperature and humidity |
-20°C to 60°C, 80% RH (no condensation) |
Vibration resistance |
JIS D 1601:1995 5.3(1), Class 1A (passenger vehicle) equivalent |
LED display |
Wireless connection, measurement status, error status, AC adapter/external power, battery power, charge status |
Automatic connection function |
Available |
Number of input channels |
5 ch (Able to set voltage or strain for each channel) |
Input terminals |
Push-button type terminal block (5 terminals per channel), outfitted with terminal block cover
Set DIP switches according to measurement target. |
Measurement target |
Voltage
Strain: Strain gage
1-gage method (2-wire setup), 1-gage method (3-wire setup), 2-gage method (adjacent sides), 4-gage method
Strain gauge transducer |
Input type |
Balanced differential input, Simultaneous sampling of all channels
(non-isolated channels) |