| Accuracy: |
23 °C ± 5 °C (73 °F ± 9 °F), 80% rh or less |
| Measurable conductor diameter: |
ø 254 mm (10") max |
| Primary input current: |
5000 A AC / 500A AC (range selectable)
Effective input level: 10 % to 100 % of range |
| Secondary output voltage: |
500 mV AC f.s. |
| Accuracy (45 to 66 Hz): |
Amplitude: ±2.0 %rdg.±1.5 mV
phase angle: ±1.0°or less |
| Frequency characteristics: |
10 Hz to 20 kHz (±3dB) |
| Effect of Conductor position: |
Within ±3.0 % in any direction from the center of sensor |
| Max. usable circuit voltage: |
1000 Vrms (insulated conductor) |
| Max. allowable current: |
10000 A rms cont. (45 Hz to 66 Hz) |
| Withstand voltage: |
7.4 kV rms between clamp and circuitry(50/60 Hz, 1 minute) |
| Applicable standards: |
Safety
EN61010-2-032:1995 CAT III 1000V
EMC
EN61326-1:1997+A1:1998 |
| Operating temperature: |
0°C to 40°C, 80 %rh or less, no condensation |
| Power supply: |
LR03 alkaline batteryX4 (continuous operation max. 168 hours) or 9445 AC ADAPTER (optional) |