Hioki CM4374 Specifications |
DC Current range |
600.0 A/2000 A, Basic accuracy: ±1.3% rdg. ±0.3 A (600A range) |
AC Current range |
600.0 A/2000 A (10 Hz to 1 kHz, True RMS), Basic accuracy: ±1.3% rdg. ±0.3 A (600A) |
Crest factor |
600.0 A range: 3 or less, 2000 A range: 2.84 or less |
DC+AC Current range |
600.0 A/2000 A (10 Hz to 1 kHz, True RMS), Basic accuracy DC, 45-66 Hz: ±1.3% rdg. ±1.3 A (at 600 A) |
DC Voltage range |
600.0 mV to 1500 V (*1), 5 ranges, Basic accuracy: ±0.5% rdg. ±0.5 mV (at 600 mV) |
AC Voltage range |
6.000 V to 1000 V, 4 ranges (15 Hz to 1 kHz, True RMS), Basic accuracy 45 - 66 Hz: ±0.9% rdg. ±0.013 V (at 6 V) |
DC+AC Voltage range |
6.000 V to 1000 V, 4 ranges, Basic accuracy DC, 45 - 66 Hz: ±1.0% rdg. ±0.023 V (at 6 V) |
Resistance range |
600.0 Ω to 600.0 kΩ, 4 ranges, Basic accuracy: ±0.7% rdg. ±0.5 Ω (at 600 Ω) |
Capacitance range |
1.000 μF to 1000 μF, 4 ranges, Basic accuracy: ±1.9% rdg. ±0.005 μF (at 1 μF) |
Frequency range |
9.999 Hz to 999.9 Hz, 3 ranges, Basic accuracy: ±0.1% rdg. ±0.003 Hz (at 9.999 Hz) |
Temperature (K) |
-40.0 to 400.0 ˚C, add temperature probe accuracy to basic accuracy of ±0.5% rdg. ±3.0 ˚C |
Voltage detection |
Hi: 40 V to 600 V AC, Lo: 80 V to 600 V AC, 50/60 Hz |
Other functions |
Automatic AC/DC detection, Plus/Minus judgement function of DC A, DC V, Max/Min/Average/PEAK MAX/ PEAK MIN value display, Filter function, Display value hold, Auto hold, Back light, Auto-power save, Buzzer sounds, Zero-adjustment |
Dustproof, waterproof |
Grip: IP54 (when measuring an insulated conductor only), Jaw (the current sensor portion of the instrument)/ barrier: IP50 *Risk of electric shock from the conductor being measured increases when wet. |
Interface |
Bluetooth® 4.0LE, Supported devices: iOS, Android™ smartphone/tablet
GENNECT Cross (free downloadable app) : display measured values and waveforms, Simple logging function, Waveform monitor function, Hold save function |
Power supply |
LR03 Alkaline battery ×2, Continuous use: 45 hr (Backlight OFF, Bluetooth® OFF), 24 hr (Backlight OFF, Bluetooth® ON) |
Core jaw diameter |
φ55 mm (2.17 in), Jaw dimension: 92 mm (3.62 in) W×18 mm (0.71 in) D |
Dimensions and mass |
65 mm (2.56 in) W×250 mm (9.84 in) H×35 mm (1.38 in) D mm, 530 g (18.7 oz) |
Accessories |
TEST LEAD L9207-10, CARRYING CASE C0203, LR03 Alkaline battery ×2, Instruction Manual ×1, Precautions Concerning Use of Equipment that Emits Radio Waves ×1 |