30 mΩ Max. display |
31.000 mΩ, resolution: 1 μΩ, measurement current: 100 mA |
300 mΩ Max. display |
310.00 mΩ, resolution: 10 μΩ, measurement current: 10 mA |
3 Ω Max. display |
3.1000 Ω, resolution: 100 μΩ, measurement current: 1 mA |
30 Ω Max. display |
31.000 Ω, resolution: 1 mΩ, measurement current: 100 μA |
300 Ω Max. display |
310.00 Ω, resolution: 10 mΩ, measurement current: 10 μA |
3 kΩ Max. display |
3.1000 kΩ, resolution: 100 mΩ, measurement current: 10 μA |
Basic accuracy |
±0.5% rdg ±5 dgt(30 mΩ range or more |
Measurement frequency |
1 kHz ±0.2 Hz |
Measurement method |
AC four-terminal method |
Voltage measurement ranges |
- 6 V(Max. display: 6.00000 V, resolution: 10 μV)
- 60 V(Max. display: 60.0000 V, resolution: 100 μV)
|
Basic accuracy |
±0.01% rdg. ±3 dgt. |
Response time |
10 ms |
Sampling period |
Ω or V (60 Hz): 4 ms (EX.FAST), 12 ms (FAST), 35 ms (MEDIUM), 150 ms (SLOW)
ΩV (60 Hz): 8 ms (EX.FAST), 24 ms (FAST), 70 ms (MEDIUM), 253 ms (SLOW)
Ω or V (50 Hz): 4 ms (EX.FAST), 12 ms (FAST), 42 ms (MEDIUM), 157 ms (SLOW)
ΩV (50 Hz): 8 ms (EX.FAST), 24 ms (FAST), 84 ms (MEDIUM), 259 ms (SLOW) |
Functions |
Contact check, Zero adjustment (±1000 counts), Pulse measurement, Comparator (Hi/ IN/ Lo), Statistical calculations (Max. 30,000), Delay, Average, Panel saving/loading, Memory storage, LabVIEW® driver |
Interfaces |
LAN (TCP/IP, 10BASE-T/100BASE-TX)
RS-232C (Max. 38.4 kbps, Available as printer I/F)
EXT I/O (37-pin Handler interface)
Analog output (DC 0 V to 3.1 V) |
Power supply |
100 to 240 V AC, 50 Hz/60 Hz, 35 VA max. |
Dimensions and mass |
215 mm (8.46 in) W × 80 mm (3.15 in) H × 295 mm (11.61 in) D, 2.4 kg (84.7 oz) |