Basic Specifications |
Measurement parameters |
C (Capacitance), D (dissipation factor tan δ), Q (1/tan δ) |
Measurement frequencies |
1 kHz and 1 MHz
Accuracy: ±0.01% or better Frequency shift: 1 MHz ±1%, ±2% |
Measurement signal level |
Open-circuit terminal voltage: 500 mV or 1 V
Signal level accuracy: ±10% ±5 mV
Output resistance: Approx. 1Ω (@1 kHz in 2.2 μF and higher
ranges), approx. 20 Ω (in ranges other than the above) |
Range of measurable values |
C: 0.000 fF to 15.0000 μF
D: 0.00001 to 1.99999
Q: 0.0 to 19999.9 |
Equivalent circuit mode |
Series-Parallel Equivalent Circuit mode (Auto/Manual) |
Measurement time |
Representative value: 1.5 ms (FAST) (Actual measurement time depends on measurement configuration settings) |
Measurement speed |
FAST, NORMAL and SLOW |
Averaging function |
1 to 256 |
Low-C reject function |
Bad Contact Detection (open circuit between terminals) |
Chatter detection functionq |
Bad Contact Detection (chatter) |
Current detection circuit monitoring function |
A measurement signal fault is detected when external noise causes measurement levels to be out of tolerance for the measurement range. |
Applied voltage value monitoring function |
An error is detected when voltage monitored across a test sample is outside of the judgment standard. |
Trigger function |
Internal and external triggers can be enabled |
Trigger delay |
0 to 9.999 s |
Memory function |
Up to 1,000 measurement values can be stored in the instrument (Downloadable by GP-IB and RS-232C) |
Compensation |
Open- and short-circuit compensation, load and cable length compensation, and self calibration |
Trigger-synchronous output function |
Functions to inhibit measurement voltage except while actually measuring |
Key-lock function |
Settings can be made and cancelled by front panel keys |
BIN measurement |
C: 13 ranks, D-NG, OUT OF BINS, Absolute value setting, ∆ setting, ∆% setting |
Comparator |
C: HI, IN, LO; D(Q): HI, IN, LO
Absolute value setting, ∆ setting, ∆% setting |
Panel save and load |
70 setting configurations can be stored Load method: key operation or external I/O |
Audible beeper |
Setting the buzzer for comparator judgment results (IN or NG) to ON or OFF is possible |
Interface |
RS-232C, GP-IB and EXT I/O (included as standard) |
Printer function |
Measurement values can be printed (requires Model 9442 or 9444 cable) |
Display device |
LEDs (on/off) |
Operating temperature and humidity |
0 to 40°C, 80% rh or less (non-condensating) |
Operating temperature and humidity |
0 to 40°C, 80% rh or less (non-condensating) |
Storage temperature and humidity |
-10 to 55°C, 80% rh or less (non-condensating) |
Operating environment |
Indoors, to 2000 m ASL |
Power supply |
Selectable 100, 120, 220 or 240 V ±10%, 50/60 Hz |
Max. power consumption |
40 VA |
Withstand voltage |
Power line to ground: 1.39 kV AC for 15 s |
Backup battery life |
Approx. 6 years |
Dimensions |
Approx. 260 W × 100 H × 298 D mm (10.24" W × 3.94" H × 11.73" D) (exc. projections) |
Mass |
Approx. 4.8 kg (169.3 oz) |
Applicable standards |
EMC: EN61326, EN61000-3-2, EN61000-3-3 Safety: EN61010-1 |
Supplied accessories |
Power Cord, Spare Fuse |