Logger w/Cable |
Measurement Range |
4-20mA (w/CABLE-4-20MA) 0 to 20.1 mA
0 to 2.5 V (w/CABLE-2.5-STEREO)
0 to 5 V (w/CABLE-ADAP5)
0 to 10 V (w/ CABLE-ADAP10)
0 to 24 V (w/ CABLE-ADAP24) |
Accuracy |
4-20mA ±0.001 mA ±0.2% of reading (w/CABLE-4-20MA)
±0.1 mV ±0.1% of reading (w/CABLE-2.5-STEREO)
±0.2 mV ±0.3% of reading (w/CABLE-ADAP5)
±0.4 mV ±0.3% of reading (w/ CABLE-ADAP10)
±1.0 mV ±0.3% of reading(w/ CABLE-ADAP24) |
Resolution |
0.3 µA 4-20mA (w/CABLE-4-20MA)
40 µV (w/CABLE-2.5-STEREO)
80 µV (w/CABLE-ADAP5)
160 µV (w/ CABLE-ADAP10)
384 µV (w/ CABLE-ADAP24) |
Logger |
Operating Range Logging |
-20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing); |
Launch/Readout |
0° to 50°C (32° to 122°F) per USB specification |
Logging Rate |
1 second to 18 hours, 12 minutes, 15 seconds |
Logging Modes |
Normal, burst, or statistics |
Memory Modes |
Wrap when full or stop when full |
Start Modes |
Immediate, push button, date & time, or next interval |
Stop Modes |
When memory full, push button, or date & time |
Restart Mode |
Push button |
Time Accuracy |
±1 minute per month at 25°C (77°F), see Plot A |
Battery Life |
1 year, typical with logging rate of 1 minute and sampling interval of 15 seconds or greater |
Battery Type |
Two AAA 1.5 V alkaline batteries, user replaceable |
Memory |
4 MB (1.9 million measurements, maximum) |
Download Type |
USB 2.0 interface |
Full Memory Download Time |
Approximately 1.5 minutes |
LCD |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
Size |
10.8 x 5.41 x 2.54 cm (4.25 x 2.13 x 1 in.) |
Weight |
107.5 g (3.79 oz) |
Environmental Rating |
IP50 |
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU). |