Specifications |
|
9170 |
9171 |
9172 |
9173 |
Range (at 23 °C ambient) |
–45 °C to 140 °C
(–49 °F to 284 °F) |
–30 °C to 155 °C
(–22 °F to 311 °F) |
35 °C to 425 °C
(95 °F to 797 °F) |
50 °C to 700 °C†
(122 °F to 1292 °F) |
Display Accuracy |
± 0.1 °C full range |
± 0.1 °C: 35 °C to 100 °C
± 0.15 °C: 100 °C to 225 °C
± 0.2 °C: 225 °C to 425 °C |
± 0.2 °C: 50 °C to 425 °C
± 0.25 °C: 425 °C to 660 °C |
Stability |
± 0.005 °C full range |
± 0.005 °C: 35 °C to 100 °C
± 0.008 °C: 100 °C to 225 °C
± 0.01 °C: 225 °C to 425 °C |
± 0.005 °C: 50 °C to 100 °C
± 0.01 °C: 100 °C to 425 °C
± 0.03 °C: 425 °C to 700 °C |
Axial Uniformity (60 mm) |
± 0.1 °C at –45 °C
± 0.04 °C at –35 °C
± 0.02 °C at 0 °C
± 0.07 °C at 140 °C |
± 0.025 °C at –30 °C
± 0.02 °C at 0 °C
± 0.07 °C at 155 °C |
± 0.05 °C: 35 °C to 100 °C
± 0.1 °C: 100 °C to 225 °C
± 0.2 °C: 225 °C to 425 °C |
± 0.1 °C: 50 °C to 100 °C
± 0.25 °C: 100 °C to 425 °C
± 0.4 °C: 425 °C to 700 °C |
Radial Uniformity |
± 0.01 °C full range |
± 0.01 °C: 35 °C to 100 °C
± 0.02 °C: 100 °C to 225 °C
± 0.025 °C: 225 °C to 425 °C |
± 0.01 °C: 50 °C to 100 °C
± 0.025 °C: 100 °C to 425 °C
± 0.04 °C: 425 °C to 700 °C |
Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes) |
± 0.02 °C at –45 °C
± 0.005 °C at –35 °C
± 0.01 °C at 140 °C |
± 0.005 °C at –30 °C
± 0.005 °C at 0 °C
± 0.01 °C at 155 °C |
± 0.01 °C full range |
± 0.02 °C at 425 °C
± 0.04 °C at 700 °C |
Hysteresis |
0.025 °C |
0.04 °C |
0.07 °C |
Well Depth |
160 mm (6.3 in) |
203 mm (8 in) |
Resolution |
0.001 °C |
Display |
LCD, °C or °F, user-selectable |
Key Pad |
Ten key with decimal and +/- button. Function keys, menu key, and °C / °F key. |
Cooling Time |
44 min: 23 °C to –45 °C
19 min: 23 °C to –30 °C
19 min: 140 °C to 23 °C |
30 min: 23 °C to –30 °C
25 min: 155 °C to 23 °C |
220 min: 425 °C to 35 °C
100 min: 425 °C to 100 °C |
235 min: 700 °C to 50 °C
153 min: 700 °C to 100 °C |
Heating Time |
32 min: 23 °C to 140 °C
45 min: –45 °C to 140 °C |
44 min: 23 °C to 155 °C
56 min: –30 °C to 155 °C |
27 min: 35 °C to 425 °C |
46 min: 50 °C to 700 °C |
Size (HxWxD) |
366 x 203 x 323 mm (14.4 x 8 x 12.7 in) |
Weight |
14.2 kg (31.5 lb) |
15 kg (33 lb) |
13.2 kg (29 lb) |
15 kg (33 lb) |
Power |
115 V ac (± 10 %), or
230 V ac (± 10 %), 50/60 Hz, 550 W |
115 V ac (± 10 %), or
230 V ac (± 10 %), 50/60 Hz, 1025 W |
Computer Interface |
RS-232 Interface with 9930 Interface-it control software included |
Traceable Calibration (NIST) |
Data at –45 °C, 0 °C, 50 °C, 100 °C, and 140 °C |
Data at –30 °C, 0 °C, 50 °C, 100 °C, and 155 °C |
Data at 100 °C, 150 °C, 250 °C, 350 °C, and 425 °C |
Data at 100 °C, 200 °C, 350 °C, 500 °C, and 660 °C |
†Calibrated to 660 °C; reference thermometer recommended at higher temperatures. |
Specifications |
Built-in Reference Input |
Temperature Range |
–200 °C to 962 °C (–328 °F to 1764 °F) |
Resistance Range |
0 Ω to 400 Ω, auto-ranging |
Characterizations |
ITS-90 subranges 4, 6, 7, 8, 9, 10, and 11 Callendar-Van Dusen (CVD): R0, a, b, d |
Resistance Accuracy |
0 Ω to 20 Ω: 0.0005 W
20 Ω to 400 Ω: 25 ppm |
Temperature Accuracy (does not include probe uncertainty) |
10 Ω PRTs:
± 0.013 °C at 0 °C
± 0.014 °C at 155 °C
± 0.019 °C at 425 °C
± 0.028 °C at 700 °C |
25 Ω and 100 Ω PRTs:
± 0.005 °C at –100 °C
± 0.007 °C at 0 °C
± 0.011 °C at 155 °C
± 0.013 °C at 225 °C
± 0.019 °C at 425 °C
± 0.027 °C at 661 °C |
Resistance Resolution |
0 Ω to 20 Ω: 0.0001 Ω
20 Ω to 400 Ω: 0.001 Ω |
Measurement Period |
1 second |
Probe Connection |
4-wire with shield, 5-pin DIN connector |
Calibration |
NVLAP accredited (built-in reference input only), NIST-traceable calibration provided |