Specifications |
9100S |
9102S |
Range |
35 °C to 375 °C (95 °F to 707 °F |
–10 °C to 122 °C (14 °F to 252 °F) at23 °C ambient |
Accuracy |
± 0.25 °C at 50 °C; ± 0.25 °C at 100 °C; ± 0.5 °C at 375 °C |
± 0.25 °C |
Stability |
± 0.07 °C at 50 °C; ± 0.1 °C at 100 °C; ± 0.3 °C at 375 °C |
± 0.05 °C |
Well-to-Well Uniformity |
± 0.2 °C with sensors of similar size at equal depths within wells |
Heating Times |
35 °C to 375 °C: 9.5 minutes |
ambient to 100 °C: 10 minutes |
Stabilization |
5 minutes |
7 minutes |
Cooling Times |
375 °C to 100 °C: 14 minutes |
ambient to 0 °C: 10 minutes |
Well Depth |
102 mm (4 in);1.6 mm (1/16 in) hole is 89 mm (3.5 in) deep |
102 mm (4 in) |
Removable Inserts |
N/A |
Available in sizes from 1.6 mm (1/16 in) to 11.1 mm (7/16 in) [6.4 mm (1/4 in) and 4.8 mm (3/16 in) included] |
Power |
115 V ac (± 10 %), 55–65 Hz, 1.5 A or 230 V ac (± 10 %), 0.8 A, 45–55 Hz, 175 W |
94-234 V ac (± 10 %), 50/60 Hz, 60 W;or 12 VDC |
Size (HxWxD) |
57 x 125 x 150 mm (2.25 x 4.9 x 6.1 in) |
99 x 140 x 175 mm (3.9 x 5.5 x 6.9 in) |
Weight |
1 kg (2 lb 3 oz) |
1.8 kg (4 lb) |
Computer Interface |
RS-232 included with free Interface-it software |
NIST-TraceableCalibration |
Data at 50 °C, 100 °C, 150 °C, 200 °C,250 °C, 300 °C, and 375 °C |
Data at –10 °C, 0 °C, 25 °C, 50 °C, 75 °C,100 °C, and 122 °C |