Measurement Accuracy |
717 Series
Measure Pressure |
(internal sensor) |
Accuracy: |
0.05% full scale |
Functions: |
Zero, Min, Max, Hold, Damp
Note Gasses / liquids (non-corrosive) |
Fluke 717 1G
Range: |
-1 to 1 PSI, (-68.9 mbar to 68.9 mbar, -83 to 6.89 kPa) |
Resolution: |
0.0001 PSI, 0.001 mbar, 0.0001 kPa |
Over Pressure: |
5 X Full Scale |
Fluke 717 30G
Range: |
-12 to 30 PSI, (-850 mbar to 2 bar, -83 to 206.84 kPa ) |
Resolution: |
0.001 PSI, 0.1 mbar, 0.01 kPa |
Over Pressure: |
2 X full scale |
Fluke 717 100G
Range: |
-12 to 100 PSI, (-850 mbar to 7 bar, -83 to 689.48 kPa) |
Resolution: |
0.01 psi, 0.1 mbar, 0.01 kPa |
Over Pressure: |
2 X full scale |
Fluke 717 300G
Range: |
-12 to 300 PSI, (-850 mbar to 20 bar, -83 to 2068.42 kPa ) |
Resolution: |
0.01 PSI, 1 mbar, 0.1 kPa |
Over Pressure: |
375 PSI, 25 bar |
Fluke 717 500G
Range: |
0 to 1000 PSI, (0 mbar to 69 bar, 0 to 6894.76 kPa ) |
Resolution: |
0.1 PSI, 1 mbar, 0.1 kPa |
Over Pressure: |
2 X full scale |
Fluke 717 1500G
Range: |
0 to 1500 PSI, (0 bar to 103.4 bar, 0 to 10342 kPa ) |
Resolution: |
0.1 PSI, 0.01 mbar, 1 kPa |
Over Pressure: |
2 X full scale |
Fluke 717 3000G
Range: |
0 to 3000 PSI, (0 bar to 207 bar, 0 to 20684.27 kPa) |
Resolution: |
0.1 PSI, 0.01 bar, 1 kPa |
Over Pressure: |
2 X full scale |
Fluke 717 5000G
Range: |
0 to 5000 PSI, (0 bar to 345 bar, 0 to 34474 kPa) |
Resolution: |
0.1 PSI, 0.01 bar, 1 kPa |
Over Pressure: |
2 X full scale |
Measure Pressure |
(via pressure module connector using optional Pressure Modules) |
Range: |
29 pressure modules, 0-1 inH20 to 10,000 PSI, 2.5 mbar to 700 bar ranges. Over pressure per pressure module specs |
Resolution: |
Per pressure module specs |
Accuracy: |
per pressure module specs |
Functions: |
Zero, Min, Max, Hold, Damp |
Range: |
0 to 24 mA (-25% to 125%) |
Resolution: |
0.001 mA |
Accuracy: |
0.015% + 2 counts |
Specifications |
Range: |
24 V dc |
Accuracy: |
+/- 10%
Note: Drive; 20 mA into 1000 ohms for battery >6.8V; 700 W for battery 5.8 to 6.8 V |
Display |
Clear 5 digit LCD |
Environmental Specifications |
Operating Temperature |
-10 ºC to 55 ºC |
Storage Temperature |
-40 ºC to 60 ºC |
Humidity (Without Condensation) |
95% (10 ºC to 30 ºC)
75% (30 ºC to 40 ºC)
45% (40 ºC to 50 ºC)
35% (50 ºC to 55 ºC) |
Operating Altitude |
3,000 M max |
Safety Specifications |
Safety |
CSA C22.2 No. 1010.1:1992 |
Mechanical & General Specifications |
Size |
201 mm L x 98 mm W x 52 mm (7.93in. L x 3.86in. W X 2.06in.) |
Weight |
22 oz, approximately 600G with holster |
Warranty |
3 years |
Shock & Vibration |
Random, 2G, 5-500 Hz
1m drop test |
Battery type |
Single 9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline |