Infrared Measurements |
Infrared temperature range |
-30 °C to 900 °C
(-22 °F to 1652 °F) |
IR accuracy (calibration geometry with ambient temperature 23 °C ± 2 °C) |
≥ 0°C: ± 1°C or ± 1 % of the reading, whichever is greater
≥-10 °C to <0 °C: ±2°C
<-10°C: ±3 °C
≥32°F: ± 2°F or ± 1% of the reading, whichever is greater
≥14 °F to <32 °F: ±4 °F
<14 °F: ±6 °F |
IR repeatability |
±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater |
Display resolution |
0.1 °C / 0.1 °F |
Distance Spot |
60:1 (calculated at 90 % energy) |
Minimum spot size |
19 mm |
Laser sighting |
Offset dual laser, output <1 mW |
Spectral response |
8 μm to 14 μm |
Response Time (95 %) |
<500 ms |
Emissivity |
Digitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of common materials |
Contact Measurements |
K-Type Thermocouple input temperature range |
-270 °C to 1372 °C (-454 °F to 2501 °F) |
K-Type Thermocouple input accuracy (with ambient temperature 23 °C ± 2 °C) |
<-40 °C: ±(1 °C + 0.2 °/1 °C)
≥-40 °C: ±1 % or 1 °C, whichever is greater
<-40 °F: ±(2 °F + 0.2 °/1 °F)
≥-40 °F: ±1 % or 2 °F, whichever is greater |
K-Type Thermocouple |
0.1 °C/0.1 °F |
K-Type Thermocouple repeatability |
±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater |
Measurement Options |
Hi/Low alarms |
Audible and two-color visual |
Min/Max/Avg/Dif |
Yes |
Switchable Celsius and Fahrenheit |
Yes |
Backlight |
Two levels, normal and extra bright for darker environments |
Probe input |
K-Type Thermocouple Simultaneous display of probe and IR temperature |
Trigger lock |
Yes |
Data storage |
99 points |
Display |
Dot matrix 98 x 96 pixels with function menus |
Communication |
USB 2.0 |
K-Type Thermocouple Specifications |
Measurement range (bead probe) |
-40 °C to 260 °C (-40 °F to 500 °F) |
Accuracy |
±1.1 °C (±2.0 °F) from 0 °C to 260 °C (32 °F to 500 °F). Typically within 1.1 °C
(2.0 °F) from -40 °C to 0 °C (-40 °F to 32 °F) |
Cable length |
1 m (40 in) K-type thermocouple cable with standard miniature thermocouple
connector and bead termination |
General Specifications |
Safety and Compliance |
IEC 60825-1
FDA Laser Class II
IEC 61326-1
CE Complaint
CMC # 01120009 |
Operating temperature |
0 °C to 50 °C (32 °F to 122 °F) |
Storage temperature |
-20 °C to 60 °C (-4 °F to 140 °F) |
Relative humidity |
10 % to 90 % RH non-condensing up to 30 °C (86 °F) |
Operating altitude |
2000 meters above mean sea level |
Weight |
0.322 kg (0.7099 lb) |
Power |
2 AA batteries |
Battery life |
8 hours with laser and backlight on; 100 hours with laser and backlight off, at
100 % duty cycle (thermometer continuously on) |