Current Ranges |
60 A / 600 A / 6000 A AC RMS |
Output Sensitivity (AC
coupled) |
50 mV /5mV / 0.5mV per A |
Load Impedance |
100 kOhm minimum |
Accuracy (at +25°C) |
±1% of range (45 – 65 Hz) |
Linearity (10% to 100% of
range) |
±0.2% of reading |
Noise |
8mV AC RMS (60 A)
2 mV AC RMS (600/6000A) |
Frequency Range |
10 Hz to 50 kHz (- 3 dB) |
Phase Error |
< ±1º (45 – 65Hz),
±10º (at 20 kHz) |
Position Sensitivity (with
cable > 25mm from the
coupling) |
±2% of range |
External field (with cable >
200mm from the head) |
±1% of range |
Power Supply |
2 x AA MN 1500 LR6 alkaline
400 Hours, low battery indicator or dedicated ext power supply |
Temperature Coefficient |
±0.08% of reading per ºC |
Working Voltage (see
Safety Standards section) |
600 V AC RMS or DC |
Head Cable length…
(double insulated) |
610mm (24”) I6000s Flex-24
915mm (36”) I6000s Flex-36 |
Head Cable Diameter |
14.3 mm |
Head Bend Radius |
38.1 mm |
Cable length (head to
electronics) |
2 meters |
Output Connection |
0.5 m cable terminated with
safety BNC connector
Supplied with 4mm safety plug
adaptor |
Operating Temperature
Range |
-20 to +90 °C (head)
-20 to +85 °C (electronics) |
Storage Temperature
Range |
-40 to +105 °C (head)
-20 to +85 °C (electronics) |
Operating Humidity |
15% to 85% (non condensing) |
Weight |
180 g (head), 190 g (electronics) |
BS EN 61010 |
1: 2001 |
BS EN 61010 |
2-032: 2002 |
BS EN 61010 |
031: 2002 |
600V RMS, Category III,
Pollution Degree 2 |
|
Note |
Use of the probe on uninsulated conductors is limited to 600 V ACRMS or DC
and frequencies below 1 kHz. |
EMC Standards |
BS EN 61326: 1998 +A1, A2, & A3 |