YSI TruLine ISE Specifications |
Ammonia (NH3)* |
400365 |
Gas- sensing |
Replaceable/refillable membrane modules (1 replacement included), pH sensor inside |
0.02 to 17,000 mg/L |
54 to 59 mV/ decade |
ISA buffers sample to
>pH 11 to make NH3 gaseous |
Few, volatile amines |
Wastewater, aqua-culture, agriculture, industrial. pH shift caused by ammonia gas is correlated to conc. |
Ammonium (NH +)
4 |
400371 |
Polymer/ PVC mem- brane |
Refillable reference, re- placeable sensor modules (1 replacement included) |
0.02 to 1,800 mg/L |
54 to 59 mV/ decade |
4 to 10 |
Few, Na+, K+ |
Wastewater, environmental, industrial, food/ beverage |
Bromide (Br-) |
400377 |
Solid state |
Refillable reference, non- replaceable sensor |
0.4 to 79,900 mg/L |
54 to 59 mV/ decade |
<12 |
I-, Cl-, CN-, NH3 -
|
Pool, spa, industrial, wastewater |
Cadmium (Cd2+) |
400382 |
Solid state |
Refillable reference, non- replaceable sensor |
0.1 to 11,200 mg/L |
26 to 29 mV/ decade |
2 to 8 |
Hg2+, Ag+, Cu2+, Pb2+ |
Industrial, environmental |
Calcium (Ca2+) |
400387 |
Polymer/ PVC mem- brane |
Refillable reference, re- placeable sensor modules (1 replacement included) |
0.02 to 40,000 mg/L |
25 to 29 mV/ decade |
2.5 to 11 |
Pb2+, Hg2+, Si2+, Mg2+, Cu2+, K+,
Tris+, Na+, others |
Aquariums, industrial, food/beverage |
Chloride (Cl-) |
400393 |
Solid state |
Refillable, reference non- replaceable sensor |
1.8 to 35,500 mg/L |
54 to 59 mV/ decade |
2 to 12 |
CN-, Br-, I-, OH-, S2-, NH3 |
Industrial, wastewater, drinking water, food/ beverage, environmental. Store in dark when not in use. |
Copper (Cu2+) |
400445 |
Solid state |
Refillable reference, non- replaceable sensor |
0.06 to 6,400 mg/L |
26 to 20 mV/ decade |
2 to 6 |
Hg2+, Ag+, Fe2+, Br-, Cl- |
Industrial, environmen- tal. Also known as Cupric ISE. |
Cyanide (CN-)** |
400403 |
Solid state |
Refillable reference, non- replaceable sensor |
0.2 to 260 mg/L |
54 to 59 mV/ decade |
10 to 14 |
I-, Br-, Cl-, S2- |
Industrial, environmental |
Fluoride (F-) |
400406 |
Solid state crystal |
Refillable reference, non- replaceable sensor |
0.02 mg/L - satura- tion |
54 to 59 mV/ decade |
5 to 7 at low
range, 11 at higher range |
OH- |
Drinking water, indus- trial. Lanthanum fluoride crystal is the sensor. |
Iodide (I-) |
400411 |
Solid state |
Refillable reference, non- replaceable sensor |
0.006 to 127,000
mg/L |
54 to 59 mV/ decade |
<14 |
CN-, Cl-, S2 O32-,S2-, NH3 |
Drinking water, industrial, food/beverage |
Lead (Pb2+) |
400416 |
Solid state |
Refillable reference, non- replaceable sensor |
0.02 to 20,700 mg/L |
26 to 29 mV/ decade |
4 to 7 |
Hg2+, Ag+, Cu2+, Fe2+, Cd2+ |
Industrial, environmental |
Nitrate (NO -)
3 |
400421 |
Polymer/ PVC mem- brane |
Refillable reference, re- placeable sensor modules (1 replacement included) |
0.4 to 62,000 mg/L |
54 to 59 mV/ decade |
2.5 to 11 |
Few, ClO4-, ClO3-, I-, F- |
Environmental, industrial, agriculture, food/ beverage |
Potassium (K+) |
400427 |
Polymer/ PVC mem- brane |
Refillable reference, re- placeable sensor modules (1 replacement included) |
0.04 to 39,000 mg/L |
54 to 59 mV/ decade |
2 to 12 |
Cs+, NH +, TI+, H+,
4
Ag+, Tris+, Na+, Li+ |
Agriculture, food/beverage, environmental, industrial |
Silver/Sulfide (Ag+/S2-)** |
400433 |
Solid state |
Refillable reference, non- replaceable sensor |
0.01 to 107,900
mg/L (Silver); 0.003
to 32,100 mg/L (Sulfide) |
54 to 59 mV/decade (Silver); 26 to
29 mV/decade (Sulfide) |
2 to 12 |
Hg2+ |
Industrial, environmental, food/beverage |
Sodium (Na+) |
400439 |
Glass sensor |
Refillable reference, non- replaceable sensor |
0.1 to 23,000 mg/L |
54 to 59 mV/ decade |
>9 |
H+, K+ |
Industrial, agriculture, power, food/beverage |