| YSI CR 4200 General Specifications |
| Reactor type |
Dry temperature control device with safety cover |
| Cell shafts |
2 x 12 cell shafts for reaction cells 16 ± 0.2 mm |
| Reaction time setting |
20 min, 30 min, 60 min, 120 min (via fixed programs)
8 freely adjustable programs: 0...180 min |
| Temperature setting |
100 °C, 120 °C, 148 °C, 150 °C via fixed programs and 8 freely adjustable programs: Room temperature ...170 °C |
| Controlling accuracy |
± 1 °C ± 1 Digit |
| Temperature stability |
± 0.5 K |
| Overtemperature protection |
190 °C ± 5 °C |
| Heating time (with empty thermoblock) from 25 °C to |
100 °C approx. 5 min
120 °C approx. 7 min
148 °C approx. 10 min |
| Temperature of the enclose at an environmental temperature of 25 °C |
< 30 °C with a block temperature of 148 °C |
| Output |
unidirectional RS232 interface with 2 banana sockets for:
• external temperature probe
or
• PC cable |
| Power supply |
230 VAC 50 Hz ± 15 %
115 VAC 60 Hz ± 15 %
Power consumption: 560 W
Fuses 2 x 6.3 AT |
| Enclosure |
PC ABS, recyclable, high temperature resistant |
| Protective class |
I according to DIN VDE 0700 part 1/11.90 |
| Insulation group |
Insulation group: B according to DIN VDE 0110/11.72 |
| Overvoltage category |
II |
| Protection |
IP 20 according to DIN 40050 |
| Ambient temperature |
Storage -25 °C to +65 °C
Operation +5 °C to +40 °C |
| Climatic class |
2 according to VDI/VDE 3540
Relative humidity:
Yearly mean: < 75 %
30 days /year: 95 %
Other days: 85 %
Light dew: yes |
| EMC |
EN61326
FCC Class A |
| Test certificates |
cETLus, CE |
| Dimensions D x W x H: |
312 x 255 x 185 mm |
| Weight |
4 kg |
| Safety standards |
EN61010
UL3101
CAN/CSA C22.2-1010
EN61010-2-010
IEC-CAN/CSA C22.2-
1010.2.010 |