YSI CR 3200 General Specifications |
Reactor type |
Dry temperature control device with safety cover |
Cell shafts |
2 x 12 cell shafts for reaction cells 16 ± 0.2 mm |
Reaction time setting |
20 min, 30 min, 60 min, 120 min (via fixed programs)
8 freely adjustable programs: 0...180 min |
Temperature setting |
100 °C, 120 °C, 148 °C, 150 °C via fixed programs and 8 freely adjustable programs: Room temperature ...170 °C |
Controlling accuracy |
± 1 °C ± 1 Digit |
Temperature stability |
± 0.5 K |
Overtemperature protection |
190 °C ± 5 °C |
Heating time (with empty thermoblock) from 25 °C to |
100 °C approx. 5 min
120 °C approx. 7 min
148 °C approx. 10 min |
Temperature of the enclose at an environmental temperature of 25 °C |
< 30 °C with a block temperature of 148 °C |
Output |
unidirectional RS232 interface with 2 banana sockets for:
• external temperature probe
or
• PC cable |
Power supply |
230 VAC 50 Hz ± 15 %
115 VAC 60 Hz ± 15 %
Power consumption: 560 W
Fuses 2 x 6.3 AT |
Enclosure |
PC ABS, recyclable, high temperature resistant |
Protective class |
I according to DIN VDE 0700 part 1/11.90 |
Insulation group |
Insulation group: B according to DIN VDE 0110/11.72 |
Overvoltage category |
II |
Protection |
IP 20 according to DIN 40050 |
Ambient temperature |
Storage -25 °C to +65 °C
Operation +5 °C to +40 °C |
Climatic class |
2 according to VDI/VDE 3540
Relative humidity:
Yearly mean: < 75 %
30 days /year: 95 %
Other days: 85 %
Light dew: yes |
EMC |
EN61326
FCC Class A |
Test certificates |
cETLus, CE |
Dimensions D x W x H: |
312 x 255 x 185 mm |
Weight |
4 kg |
Safety standards |
EN61010
UL3101
CAN/CSA C22.2-1010
EN61010-2-010
IEC-CAN/CSA C22.2-
1010.2.010 |