Specifications |
Max Voltage |
(Between any two connections or ground)
HV7: 7KVDC 5KVAC, HV10: 10KVDC 7KVAC, HV15: 15KVDC 10KVAC |
Frequency |
<500Hz. |
Carrying Current |
<1Arms continuous, <2Arms for <1 second. |
Switching Power |
HV7 and HV10: <50W (resistive). HV15: <10W (resistive). |
Switching Time |
<5ms (including bounce). |
Expected Life |
<1mA, <100V resistive 500,000 operations. At max switching power: 1,000 operations |
Contact Resistance |
At terminals, <50% expected life operations (add 0.1Ω for <100% life)
<0.25Ω (uncompensated), <0.1Ω (compensated), <±0.1Ω difference between lowest and highest. |
Leakage Resistance |
Any individual connection to ground: >500GΩ at <30C. Common connection to ground: >100GΩ at <30C. Between any two connections: >1TΩ at <30C. |
Leakage Capacitance |
Any individual connection to ground: 15pF (typical). Common connection to ground: 50pF (typical). Between any two connections: 5pF (typical). |
Thermal EMF |
<1mV. |
Coil Power |
Holding closed: 0.3W (typical), Closing: 1.25W for 5ms (typical) |
LV Series Relay Boards |
Max Voltage |
(Between any two connections) 3KVDC 2KVAC
(between any connection and ground) 5KVDC 3KVAC |
Frequency |
<500Hz. |
Carrying Current |
<1Arms continuous, <8Arms for <1 second. |
Switching Power |
<500W (resistive) |
Switching Time |
<10ms (including bounce). |
Expected Life |
<1mA, <100V resistive 5.000,000 operations. At max switching power: 100,000 |
Contact Resistance |
At terminals, <50% expected life operations (add 0.05Ω for <100% life)
<0.175Ω (uncompensated), <0.05Ω (compensated), <±0.075Ω difference between lowest and highest. |
Leakage Resistance |
Any individual connection to ground: >1GΩ at <30C. Common connection to ground: >200MΩ at <30C. Between any two connections: >1GΩ at <30C. |
Leakage Capacitance |
Any individual connection to ground: 15pF (typical). Common connection to ground: 50pF (typical). Between any two connections: 10pF (typical). |
Thermal EMF |
<200uV. |
Coil Power |
Holding closed: 0.12W (typical), Closing: .5W for 5ms (typical) |
High Current Relay Specifications – |
Max Voltage |
Between any two connections: 1.5KVdc/1KVrms.
Any connection to ground: 1.5KVdc/1KVrms. |
Frequency |
<500Hz. |
Carrying Current |
<40Arms continuous, <60Arms for <1 second. |
Switching Power |
<500W (resistive). |
Min Switching Current |
>0.5Arms. |
Switching Time |
<20ms (including bounce). |
Expected Life |
1 to 10Arms resistive 500,000 operations. At max power: 10,000 operations. |
Contact Resistance |
<0.025Ω (at terminals, <100% expected life operations). |
Leakage Resistance |
Any individual connection to ground: >20MΩ at <30C. Common connection to ground: >5MΩ at <30C. Between any two connections: >20MΩ at <30C. |
Coil Power |
Holding closed: 0.6W (typical), Closing: 2.5W for 25ms (typical) |
Low Current Relay Specifications – |
Max Voltage |
Between any two connections: 1.5KVdc/1KVrms. Any connection to ground: 1.5KVdc/1KVrms. |
Frequency |
<500Hz. |
Carrying Current |
<1Arms continuous, <2Arms for <1 second. |
Switching Power |
<30W (resistive). |
Switching Time |
<5ms (including bounce). |
Expected Life |
<1mA, <100Vdc 5,000,000 operations. At max power: 100,000 operations. |
Contact Resistance |
At terminals, <50% expected life operations (add 0.05Ω for <100% life), <0.15Ω (uncompensated), <0.05Ω (compensated). |
Leakage Resistance |
Any individual connection to ground: >100GΩ at <30C. Common connection to ground: >10GΩ at <30C. Between any two connections: >100GΩ at <30C. |
Leakage Capacitance |
Any individual connection to ground: 15pF (typical). Common connection to ground: 50pF (typical). Between any two connections: 10pF (typical) |
Thermal EMF |
<100uV. |
Coil Power |
Holding closed: 0.06W (typical), Closing: 0.25W for 5ms (typical) |
General Specifications |
Relay Card Capacity |
Eight cards |
Front input Terminals |
5 inputs terminals |
Rear Panel Terminals |
60 terminals |
Interfaces |
RS232 and VICL (Vitrek Hipot Control) standard, GPIB optional |
Nominal Dimensions |
89mmH x 432mmW x 457mmD (3.5” x 17” x 18”) |
Nominal Weight |
Mainframe only 3.5Kg (8lb) net, 7Kg (16lb) shipping. Add 0.75Kg (1.5lb) per relay card |
Storage Environment |
-20 to 75C (non-condensing) |
Operating Environment |
0 to 50C, <85% RH (non-condensing), Pollution Degree 2 |
Operating Altitude |
0 to 10000ft ASL |
Line Power |
105-265Vrms, 45 to 450Hz, Category II using provided external power module |