Ground Bond Tests (GB, GBez - 952i, 954i, 959i) |
Test Current |
0.1 to 40A RMS, 0.001A resolution |
Accuracy |
0.5% + 5mA accuracy (add 0.005% per Hz above 100Hz) |
Test Frequency |
40Hz to 500Hz |
Resolution |
0.1Hz (1Hz above 99.9Hz) Accuracy: 0.1% accuracy |
Waveform |
Digitally synthesized, low distortion sinewave |
Measurement Configuration |
4-Terminal Kelvin |
Compliance Voltage |
6.5V RMS, may be user limited to a lower level with 0.01V resolution |
Resistance Range |
6.5 ohms at 1A decreasing to 162.5 milli-ohms max at 40A |
Max load impedance |
10 ohms |
Ramp Time |
0 to 9999sec, 0.01sec resolution |
Accuracy |
(1% + 0.1sec) accuracy |
Dwell Time |
0.02 to 9999sec or user terminated, 0.01sec resolution |
Accuracy |
0.05% + 20ms |
Ramp Down |
Fast (35mS), same rate as ramp up or may be skipped if next test step is a ground bond test |
Voltage Sense |
Range: 0 to 8 v rms |
Resolution |
4 digits down to 10uV Accuracy: 0.5% + 30uV |
Phase Measurement |
RMS, In-phase and Quadrature measurements 0.01° per Hz phase relative to test current |
Low Resistance Measurement |
Measurement Range |
0 to 150Kohm (999.9mΩ to 99.99KΩ, 149.9KΩ in 7 ranges) |
Resolution |
4 digits, 100µΩ on 1ohm range |
Test Current |
55mAdc constant current up to ≈ 91Ω, 5VDC constant V above |
Accuracy (4-wire) |
0.5% + 0.002ohm + ½ digit up to 30K ohm
1.5% + ½ digit above 30K ohm
5% + 1 digit from 100K to 150K ohm
Add 20mΩ for 2 -wire |
Line Leakage Current & Voltage Measurement (Models 951i - 955i only) |
Voltage Measurement |
0 to ± 8KVDC 6KV RMS AC (951i & 952i)
0 to ± 11KVDC 8KV RMS AC (953i & 954i)
0 to ± 11KVDC 10KV RMS AC ( 955i) |
Resolution |
0.1V, 1 V above 999.9V |
DC Accuracy |
0.25% + 0.5V |
AC Accuracy |
0.5% + 1.5V |
Leakage Current |
0 to ± 200mA DC or RMS AC |
Resolution |
4 digits (9999 counts) down to100pico-amps |
DC Accuracy |
0.25% + 0.5nA |
AC Accuracy |
0.5% + 20nA |
Test Results Test Time |
0.02 to 9999 sec Last, Minimum, Average & Max V& A reading plus arc current |
Pulse Mode (Flash) Test (Option PMT-1 available on models 951i, 952i, 953i, 954i) |
Test Waveform |
Trapezoidal (Selectable positive polarity, negative polarity or bi-polar) |
Ramp Up/Down Time |
1ms (0.5ms for option AC2) to 30mS with 0.1ms resolution |
Dwell Time |
1ms (0.5ms for option AC2) to 30mS with 0.1ms resolution |
Test Voltage |
50V to 8000V (20V to 2750V with option AC-2) |
Resolution |
0.1V up to 999.9V, 1V above |
Accuracy |
0.25% + 1.5V |
General Specifications |
Test Lead Safety Sense |
TLSS™ Technology continuously verifies that test leads are properly connected prior to and during HV, 4-wire Low Ohms and Ground Bond testing |
Arc Detection Sensitivity |
User selectable 3 microseconds to 30 microseconds. |
Dwell Time Accuracy |
Accuracy: 0.05% + 20mS |
RS232 Interface |
Digital I/O, VICL: Vitrek auxiliary unit control loop
Remote Interface Option "IF-3":
GPIB - IEEE-488.2 Subset compatible
RS232C - 1200, 2400, 4800, 9600 Baud rate selectable
Parallel - IBM/Centronics compatible |
Accuracy |
Stated specifications are bipolar and apply for one year within 5°C from calibration temperature (20 to 25°C). |
Operating Temperature |
0 C to 50° C |
Humidity |
90% RH, 0 to 40° C |
Power |
115/230 VAC + 10%, 350 VA Max |
Dimensions |
89mm H x 432mm W x 457mm D (3.5" H x 17" W x 18" D) |
Weight |
9Kg (18 lb.) Net / 18Kg (25 lb.) shipping (951i, 953i, 959i)
14Kg (28 lb.) Net / 18Kg (35 lb.) shipping (952, 954i, 955i) |