4700 Direct Input Measurement |
Maximum Input Voltage |
10KV DC, 10KV AC RMS, 15KV peak (110MO input impedance) |
DC Voltage Accuracy |
0.03% of reading + 0.03V + ½ digit |
AC Voltage Accuracy (True RMS, Band Selectable) |
0.01 to 1Hz : 0.12% of reading + 0.1V + ½ digit
10 to 65Hz : 0.12% of reading + 0.1V + ½ digit
30 to 200Hz : 0.12% of reading + 0.1V + ½ digit
200 to 450Hz : 0.4% of reading + 0.1V + ½ digit |
4700 HV-SmartProbe™ Input Measurement |
Maximum Input Voltages |
HVL & HVP 35KV Probes 35KVDC, 30KV AC RMS, 42.5KV peak (200MΩ input impedance)
HVL & HVP-70KV Probes 70KVDC, 50KV AC RMS, 71KV peak (400MΩ input impedance)
HVL-100 Probe 100KVDC, 70KV AC RMS, 100KV peak (600MΩ input impedance) |
DC Voltage Accuracy |
HVL-35: 0.025% of reading + 0.07V + ½ digit
HVP-35: 0.035% of reading + 0.07V + ½ digit
HVL-70: 0.03% of reading + 0.2V + ½ digit
HVP-70: 0.04% of reading + 0.2V + ½ digit
HVL-100: 0.05% of reading + 0.3V + ½ digit |
AC Voltage Accuracy (True RMS, Band Selectable) |
HVL/P-35 0.01 to 1Hz : 0.15% of reading + 0.2V + ½ digit
HVL-35 30 to 200Hz : 0.05% of reading + 0.2V + ½ digit*
HVP-35 30 to 200Hz : 0.1% of reading + 0.2V + ½ digit
HVL/P-35 200 to 450Hz : 0.6% of reading + 0.2V + ½ digit |
*measurement accuracy within ± 10Hz of calibration frequency |
General Specifications |
AC/DC Voltage Measurement Resolution |
Displayed Resolution: |
Selectable 4, 5 or 6 digits |
Direct Input |
.01V to 1400Vpk 0.1V to 14KVpk 1V above 14KVpk
35KV Probes .01 to 800Vpk 0.1V to 8KVpk 1V above 8KVpk
70KV Probes .01V to 1.6KVpk 0.1 V to 16KVpk 1V above 16KVpk
100KV Probe .01V to 2.4KVpk 0.1V to 24KVpk 1V above 24KVpk |
Environmental/ Physical |
Operating Environment |
0 to 50C, <85% RH (non-condensing) |
Dimensions |
218mm (8.5") wide x 130mm (5") high x 253mm (10")deep, nominally |
Weight |
2.4kg max (4kg shipping) |
Power |
AC Source |
45 to 450Hz, 100 to 265Vrms at <15W (20VA) using supplied external power supply |
DC Source |
11 to 16Vdc at <1.2A, using a center positive 2.5mm DC power connector |
BP-47 Battery Option |
Up to 11hrs continuous operation from fully charged |
Standard Accessories |
The 4700 is shipped with a NIST traceable certificate of calibration with data, direct input test leads, chassis ground lead, and external power supply. The HVP series handheld probes ship with a detachable probe tip and the HVL series probes ship with external power supply. |
Warranty |
3 year extended parts and labor warranty with no charge registration and annual factory calibration. 1 year standard warranty. |