Copper Testing |
Certification testing |
TIA 568-C.2 Cat 5e, 6, 6A ISO/IEC 11801, EN 50173 Class D, E, EA
Additionally Certifier40G supports TIA 568-C.2-1 Cat 8 ISO/IEC 11801, EN 50173 Class F, FA |
Autotest time - Cat6A/Class EA |
12 s |
Autotest time - Class FA |
19 s |
Maximm cable length for dualended autotest |
500 m |
Insertion cycles (typical) |
Channel: 10,000
Permanent link: 5,000 |
Test parameters |
As specified in TIA 1152, IEC 61935-1 |
Wiremap |
Loop resistance |
0 to 40 Ω, ±0.1 Ω |
Length |
0 to 500 m, ±0.5 m (dual-ended testing) |
Propagation delay and delay skew |
0 to 5000 ns, ±1 ns |
Attenuation |
0 to 70 dB, ±0.1 dB |
Pair-to-pair and power sum NEXT |
0 to 85 dB, ±0.2 dB |
Pair-to-pair and power sum ACR-F |
0 to 85 dB, ±0.2 dB |
Return loss |
0 to 40 dB, ±0.2 dB |
ACRN, PS ACRN |
-25 to 85 dB, ±0.5 dB |
Advanced diagnostics |
Time-domain fault locator for RL and NEXT (200 m) |
Measurement accuracy |
ANSI/TIA-1152 Level 2G, IEC 61935-1 Draft Level VI (Certifier40G)
ANSI/TIA-1152 Level IIIe, IEC 61935-1 Level IIIe (Certifier10G) |
Measurement frequency range |
1 to 2,500 MHz (Certifier40G)
1 to 500 MHz (Certifier10G) |
Fiber Testing - Multimode Encircled Flux Tier 1 (Loss/Length) Test Probes (Certifier40G only) |
Wavelengths |
850 nm, 1300 nm |
Autotest time |
10 seconds for dual-ended test |
Test standards |
TIA-568.3, IEC 14763-3, ISO 11801, Network Limits |
Measurement length |
0 to 1,000 m, ±1.5 m |
Transmitter |
Connector type |
FC |
EF compliance |
IEC 61280-4-1, TIA-526-14C |
Power |
-16 to -20 dBm |
Source type |
LED |
Spectral width |
±15 nm |
Receiver |
Connector type |
Interchangeable (SC, LC, ST) |
Dynamic range |
12 dB |
Sensitivity |
-40 dBm |
Resolution |
0.01 dB |
Visual fault locator |
Wavelength |
650 nm |
Power |
0 dBm |
Fiber Testing - Single-Mode Tier 1 (Loss/Length) Test Probes (Certifier40G only) |
Wavelengths |
1310 nm, 1550 nm |
Autotest time |
10 seconds for dual-ended test |
Test standards |
TIA-568.3, IEC 14763-3, ISO 11801, Network Limits |
Measurement length |
0 to 10,000 m, ±1.5 m |
Transmitter |
Connector type |
SC |
Power |
-5 dBm to -9 dBm |
Source type |
Laser |
Receiver |
Connector type |
SC (Adapters for LC available) |
Dynamic range |
31 dB |
Sensitivity |
-40 dBm |
Resolution |
0.01 dB |
General Specifications |
Interfaces |
Probe Interface, RJ45 Ethernet, USB host and device, talkset, power jack |
Display |
Industrial LCD, 6" touch sensitive area on both local and remote units |
Enclosure |
Ruggedized plastic with rubber overmold, withstands 1.5 m drop test on hard surface |
Remote unit capabilities |
View test results, edit labels, save results, start autotest |
Power supply |
AC 100 to 240 V to 12 V, 3 A power adapter |
Input over-voltage protection |
Protected against Telco voltages |
Battery |
Removable and rechargeable LiON Battery capacity >8 hours continuous operation |
Internal storage capacity |
700 MB |
Dimensions |
232 X 126 X 87 mm (9.1 X 5.0 X 3.4 in) |
Weight |
Approx. 1 kg (2.2 lb) per unit |
Calibration period |
1 year (copper calibration stored on mainframe, fiber calibration stored on fiber test probes) |
Environmental |
Operating temperature |
0 to 40°C |
Storage temperature |
-20 to 60°C |
Relative humidity |
10% to 80% |