| Relative Humidity |
| Measurement range |
0 ... 100 %RH |
| Accuracy (incl. non-linearity, hysteresis, and repeatability): |
| Temperature range |
0 °C ... +40 °C |
| 0 ... 90 %RH |
1.5 %RH |
| 90 ... 100 %RH |
± 2.5 %RH |
| temperature range |
-40 °C...0 °C and +40 °C ... +80 °C |
| 0 ... 90 %RH |
± 3.0 %RH |
| 90 ... 100 %RH |
± 4.0 %RH |
| Factory calibration uncertainty at +20°C (68°F): |
| 0 ... 90 %RH |
± 1.1 %RH |
| 90 ... 100 %RH |
± 1.8 %RH |
| Humidity sensor |
HUMICAP® 180R |
| Stability |
± 2 %RH over 2 years |
| Temperature |
| Measurement range |
-40 °C ... +80 °C |
| Accuracy over temperature range: |
| 0 ... +40 °C |
± 0.2 °C |
| -40 ... 0 °C, +40 ... +80 °C |
± 0.4 °C |
| Temperature Sensor |
Pt1000 RTD Class F0.1 IEC 60751 |
| General |
| Operating temperature range for probe |
-40 °C ... +80 °C |
| Probe weight with standard cable |
31 g |
| Probe housing material |
stainless steel |
| Probe filter and sensor protection |
membrane filter with chromecoated ABS plastic |
| Cable material |
Wire PVC / jacket PU |
| Cable connector |
TRRS male 3.5 mm |
| Probe housing classification |
IP65 |
| Borehole diameter needed |
16 mm |
Measurement depth with standard
accessories |
min. 30 mm, max. 90 mm |
| HM40 Indicator |
| Operation temperature range for indicator |
-10 ... +60 °C |
| Storage temperature range |
-30 ... +70 °C |
| Weight |
| Indicator with adapter |
240 g |
| SHM40 case with standard content |
3.7 kg |
| Indicator materials |
PC/ABS blend, acrylic display lens |
| Indicator adapter materials |
Nickel plated brass and plastic
overmolding |
| Indicator housing classification |
IP54 |
| Mechanical drop endurance |
1.0 m without the probe |
| Power-up time |
< 3 s |
| Alkaline batteries |
2 x AA sized, 1.5V (LR6) |
| Operation time (Alkaline batteries) |
typical 100 hours
(without backlight) |
| Calculated variables |
Td, Tw, a, x, h |
| Menu languages |
English, German, French, Finnish, Spanish,
Swedish, Chinese (simplified), Russian, Japanese |
| Display |
LCD (140 x 160 pixels) |
Electromagnetic
compatibility (EMC) |
European Union directive EN61326-1 for
portable equipment |