| Vaisala HMW92 Specifications |
| Measurement parameter(s) |
Relative humidity
Temperature
Absolute humidity*
Dew point depression*
Dew point temperature*
Enthalpy*
Mixing ratio*
Wet bulb temperature* |
| Measurement range |
0 ... 100 % RH
-5 ... +55 °C (+23 ... +131 °F) T |
| Measurement accuracy |
At temperature +10 ... 40 °C (+50 ... +104 °F)
and relative humidity 0 ... 90%:
±1.7 % RH
±0.3 °C T |
| Operating temperature |
-5 ... +55 °C (+23 ... +131 °F) |
| Operating voltage |
18 ... 35 VDC / 24 VAC |
| Analog outputs |
0 ... 5 V
0 ... 10 V (default) |
| Housing classification |
IP30 |
| Performance |
| Measurement range |
0 ... 100 %RH, non-condensing |
| Accuracy |
|
Temperature range
|
+10 ... +40 °C (+50 ... +104 °F)
|
|
0 ... 90 %RH
|
±1.7 %RH
|
|
90 ... 100 %RH
|
±2.5 %RH
|
|
Temperature range
|
-5 ... +10 °C, +40 ... + 55 °C
(+23 ... +50 °F, +104 ... +131°F)
|
|
0 ... 90 %RH
|
±3 %RH
|
|
90 ... 100 %RH
|
±4 %RH
|
| Stability in typical HVAC applications |
±0.5 %RH/year |
| Humidity sensor |
Vaisala HUMICAP ® 180R |
| Measurement range |
-5 ... +55 °C (+23 ... +131 ºF) |
| Accuracy |
|
+20 ... +30 °C (+68 ... +86 °F)
|
±0.2 °C (± 0.36 °F)
|
|
+10 ... +20 °C, +30 ... +40°C
(+50 ... +68 °F, +86 ... +104 °F)
|
±0.3 °C (± 0.54 °F)
|
|
-5 ... +10 °C, +40...+55°C
(+23 ... +50 °F, +104 ... +131 °F)
|
±0.5 °C (± 0.90 °F)
|
| Temperature sensor |
Digital temperature sensor |
| Operating temperature range |
-5 ... +55 °C (+23 ... +131 °F) |
| Storage temperature range |
-30 ... +60 °C (-22 ... +140 °F) |
| Electromagnetic compliance |
EN61326-1, Industrial Environment |
| IP class |
IP30 |
| Standard housing color |
White (RAL9003*) |
| Optional housing colors |
Black (RAL9005*) |
| (configurable models only) |
Grey (RAL7035*)
Light Ivory (RAL1015*) |
| Housing material |
ABS/PC, UL-V0 approved |
| Output connector |
Screw terminals
max. wire size 2 mm 2 (AWG14) |
| Service port connector |
4-pin M8 |
| Weight |
155 g |
| *RAL code is only indicative with potential small variations in color shade |
| Current output models |
|
Outputs
|
2 x 4 ... 20 mA, loop powered
|
|
Loop resistance
|
0 ... 600 Ω
|
|
Supply voltage
|
20 ... 28 VDC at 500 Ω load
10 ... 28 VDC at 0 Ω load
|
|
Isolation between output channels
|
500 VDC
|
| Voltage output models |
|
Outputs
|
2 × 0...5 V or 2 × 0...10 V
|
|
Load resistance
|
10 kΩ min
|
|
Supply voltage
|
18 ... 35 VDC, 24 VAC ±20 % 50/60 Hz
|
|
Max. current consumption
|
2 mA max. with relay 25 mA
|
|
Relay
|
1 pc (max 50VDC/50 VAC, 500 mA)
|
| Digital models |
|
Supply voltage
|
18 … 35 VDC, 24 VAC ± 20% 50/60 Hz
|
|
Max. current consumption
(with 120 O termination)
|
30 mA at 24 VDC
|
|
Output type
|
RS-485 (galvanic isolation, 1.5 kV)
|
|
RS-485 end of line termination
|
Enable with jumper, 120 Ω
|
|
Supported protocols
|
Selectable by DIP switch
|
| BACnet MS/TP |
| Operating mode |
Selectable Master/Slave |
|
Address range, master mode
|
0 … 127
|
|
Address range, slave mode
|
128…255
|
| Modbus RTU |
|
Address range
|
0 … 247
|
| Service port |
RS-485 line for temporary service use |