Environmental |
Operating Temperature |
14° to 122°F (-10° to 50°C) |
Storage Temperature |
-40° to 140°F (-40° to 60°C) |
Humidity |
Non condensing <50°F (10°C)
85% RH: 50° to 86°F (10° to 30°C)
70% RH: 86° to 104°F (30° to 40°C)
45% RH: 104° to 122°F (40° to 50°C) |
Altitude |
Operating - up to 200 m
Storage - 10000 m |
General |
Dimension |
18.3 (H) x 9.4 (W) x 4.3 (D) cm |
Weight |
Approx. 460 g (16.2 oz) |
Battery |
2 x LR03 (AAA) type 1.5V |
Certification |
CE |
Safety |
IEC 1010-1 (2001), UL 3111-1 (6, 1994),
EN 61010-1 (2001), CSA C22.2
No. 1010.1 (1992) |
CAT I |
OVERVOLTAGE (Installation) CATEGORY I,
Pollution Degree 2 per IEC 1010-1 |
Immersion (30 min)
& Dust proof |
IP67 according to IEC 60529 |
TP6K K-Type Bead Thermocouple (a standard accessory) |
Temperature range |
-40˚ to +900˚F (-40˚ to +480˚C) |
Type |
K-Type, Chromel Alumel, bead style |
Accuracy |
±2.0˚F (±1.1˚C) |
Electrical |
Measurement range |
J-Type: -346° to +2192°F (-210° to +1200°C)
K-Type: -328° to +2498°F (-200° to +1370°C)
T-Type: -418° to +752°F (-250° to +400°C)
E-Type: -238° to +1832°F (-150° to +1000°C) |
Display Resolution |
0.1°F/°C < 1000°
1.0°F/°C ³ 1000° |
Measurement Accuracy |
J, K, T, and E-Type; ±[0.1% +1.0°F (0.5°C)]
[Below -148°F (-100°C): add 0.2% of reading for J, K, and E-Type; and 0.1% of reading for T-Type] |
Temperature
Coefficient |
0.01% of reading 0.1°F per °F(+0.05°C per °C)
For <+64°F (+18°C) or +82°F (+28°C)
[Below -148°F (-100°C): add 0.05% of reading for J, K, and E-type; and 0.1% of reading for T-type] |
Real Time
Clock Tolerance |
About 1 second per day |
Maximum Differential
Common Mode
Voltage |
1V (maximum voltage difference between any pair of inputs) |
Temperature Scale |
ITS-90 (International Temperature Scale of 1990) |
Applicable Standards |
N.I.S.T. Monograph 175 revised to ITS-90 |
Accuracy is specified for ambient temperatures between 64°F (18°C) and 82°F (28°C) for a period of 1 year. The above specifications do not include thermocouple error. |