Accuracy (Photometric) |
±0.002A at 0.5A |
Certifications/Compliance |
CE / TUV |
Detector Type |
Dual Silicon Photodiodes |
Display |
7-inch color touchscreen, fixed, high definition, 800 × 1280 pixels |
Photometric Accuracy Instrument |
±0.002A at 0.5A±0.004A at 1.0A±0.008A at 2.0A |
Printer |
Snap-on printer included |
Wavelength Accuracy |
±0.5 nm |
Applications |
Water and Waste Water Analysis |
Baseline Flatness |
±0.002A |
Compartments |
Accessible from the top, front, or side Able to accommodate cells up to 100 mm cells Available accessory for test tubes up to 25 mm diameter and 150 mm height Removable, washable sample compartment liner with magnetic placement and hold-down |
Connections |
Single USB-A supports flash memory devices for method and data storage expert data to network or PC via USB, or EthernetPrint via USB, Ethernet or Wi-Fi USB adaptor available |
Depth (Metric) |
35.5 cm |
Description |
AquaMate 8100 UV-Vis Spectrophotometer with North America, European, and UK Power cord |
Dimensions (L x W x H) |
35.5 x 38.5 x 19.5 cm |
Drift |
<0.0005A/Hr (At 500 nm after 1-hour warmup) |
Electrical Requirements |
External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100- 240 volts, 50 -60 Hz |
Height (Metric) |
19.5 cm |
Interface |
Glove-friendly touch-screen |
Keypad |
Glove-friendly touchscreen |
Lamp |
Xenon Flash Lamp (>5 years typical) |
Length (Metric) |
35.5 cm |
Noise |
(RMS at 500 nm 60 consecutive measurements) ≤0.00020A at 0A at 260 and 500 nm ≤0.00030A at 1A at 260 and 500 nm ≤0.00040A at 2A at 260 and 500 nm |
Optical Design |
Dual-beam |
Photometric Display |
-3A to +5A |
Photometric Range |
-2A to +3.5A |
Photometric Repeatability |
±0.001A at 1A measured at 1.0A at 546 nm |
Sample Volume (Metric) |
Multiple |
Scan Speed |
Slow, medium, and fast - up to 1600 nm/min |
Spectral Bandwidth |
2 nm |
Stray Light |
< 1.0%T 198 nm (KCl) , <0.05%T at 220 nm (NaI), <0.03%T at 340 nm (NaNO2) |
Type |
Uv-Vis |
Wavelength Data Interval |
0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm |
Wavelength Range |
190 nm 1100 |
Wavelength Repeatability |
< ±0.2 nm |
Wavelength Scan Speed |
Slow, medium, and fast - up to 1600 nm/min |
Weight (Metric) |
7.5 kg |
Width (Metric) |
38.5 cm |
Warranty |
1 Year |
Unit Size |
Each |