Parameters
Differential pressure |
Measuring range |
0 to 10 Pa
0 to 50 Pa
0 to 100 Pa
0 to 500 Pa
0 to 10 hPa |
-10 to +10 Pa
-50 to +50 Pa
-100 to +100 Pa
-500 to +500 Pa
-10 to +10 hPa |
Measurement uncertainty |
±0.3% of measurement range final value ±0.3 Pa
Temperature gain drift: 0.02% of measuring range per Kelvin deviaton from nominal temperature 22 °C
Zero point drift: 0% (thanks to cyclic zeropoint adjustment) |
Selectable units |
Differential pressure in Pa, hPa, kPa, mbar, bar, mmH2O, kg/cm2, PSI, inch HG, inch H2O |
Sensor |
Piezoresistive sensor |
Autom. zero-point adjustment |
via magnetic valve
Frequency adjustable: 15 sec, 30 sec, 1 min, 5 min, 10 min |
Overload |
Measuring range Overload
0 to 10 Pa 20000 Pa
0 to 50 Pa 20000 Pa
0 to 100 Pa 20000 Pa
0 to 500 Pa 20000 Pa
0 to 10 hPa 200 hPa
-10 to 10 Pa 20000 Pa
-50 to 50 Pa 20000 Pa
-100 to 100 Pa 20000 Pa
-500 to 500 Pa 20000 Pa
-10 to 10 hPa 200 hPa |
Parameters
Humidity/temperature optional |
Probe |
Integrated probe
|
testo 6613 |
testo 6614 |
testo 6615 |
testo 6617 |
Type |
Channel |
Duct heated |
Cable trace humidity |
Cable with cover electrode monitoring |
Parameters |
%RH / °C/°F / °Ctd / °Ftd / g/kg / gr/lb / g/m3 / gr/ft³ / ppmV / °Cwb / °Fwb / kJ/kg / mbar / inch H2O / °Ctm (H2O2)/°Ftm (H2O2) / % Vol |
Meas. range |
Humidity / trace humidity |
0 to 100 %RH |
-60 to +30 °C td |
0 to 100 %RH |
Temperature |
-20 to +70 °C
-4 to +158 °F |
-40 to +180 °C
-40 to +356 °F |
40 to +120 °C
-40 to +248 °F |
-40 to +180 °C
-40 to +356 °F |
Measurement uncertainty |
Humidity |
Integrated probe |
testo 6613 |
testo 6614 |
testo 6615 |
testo 6617 |
±(1.0 + 0.007 * MV) %RH
for 0 to 90 %RH
±(1.4 + 0.007 * MV) %RH
for 90 to 100 %RH |
±(1.0 + 0.007*MV)
%RH for 0 to100 %RH |
|
±(1.2 + 0.007*MV)
%RH for 0 to 90 %RH
±(1.6 + 0.007*MV)
%RH for 90 to 100 %RH |
for deviations from media temp. ±25 °C:±0.02 %RH/K |
Dewpoint |
|
±1 K at 0 °Ctd
±2 K at -40°Ctd
±4 K at -50 °Ctd |
|
Temp. at +25°C / +77°F |
±0.15 °C / 32.2 °F
Pt1000 Class AA |
±0.15 °C/ 32.2 °F
Pt100 Class AA |
±0.15 °C/ 32.2 °F
Pt1000 Class AA |
Inputs/outputs
Analog outputs |
Quantity |
Standard: 1;
with optional humidity probe: 3 |
Output type |
0/4 to 20 mA (4-wire) (24 VAC/DC)
0 to 1/5 to 10 V (4-wire) (24 VAC/DC) |
Scaling |
Differential pressure: scalable ±50% of measuring range final value; freely scalable within measuring range |
Meas. cycle |
1/sec |
Resolution |
12 bit |
Max. load |
max. 500 Ω |
Other outputs |
Ethernet |
Optional Relay :Optional: 4 relays (free allocation to measurement channels or as collective alarm in operating menu/P2A), up to 250 VAC/3A (NO or NC) |
Digital |
Mini-DIN for P2A software |
Voltage supply |
20 to 30 VAC/DC, 300 mA current consumption, galvanically separate signal and supply line |
General technical data |
Model |
Material |
Front plate stainless steel, housing plastic |
Dimensions |
without humidity/temperature: 246 x 161 x 47 mm
with humidity/temperature:396 x 161 x 78 mm |
Weight |
Version without humidity: 0.9 kg;
Version with integrated humidity
probe:1.35 kg; version with preparation
for external humidity probe: 1.26 kg |
Display |
Display |
optional: 3-line LCD with multi-language operating menu |
Resolution |
Differential pressure |
Measuring range
0 to 10 Pa
0 to 50 Pa
0 to 100 Pa
0 to 500 Pa
0 to 10 hPa
-10 to 10 Pa
-50 to 50 Pa
-100 to 100 Pa
-500 to 500 Pa
-10 to 10 hPa |
Resolution
0.1 Pa
0.1 Pa
0.1 Pa
0.1 Pa
0.01 hPa
0.1 Pa
0.1 Pa
0.1 Pa
0.1 Pa
0.01 hPa |
Humidity |
0.1 %RH |
Temperature |
0.01 °C / 0.01 °F |
Miscellaneous |
Protection class |
IP 65 |
Connection nipple |
Ø 6 mm --> suitable hoses 4 mm + 4.8 mm |
Operating conditions |
With / without display |
Operation temperature |
-5 to +50 °C / +23 to +122 °F |
Storage temperature |
-20 to +60 °C / -4 to +140 °F |
Process temperature |
-20 to +65 °C / -4 to +149 °F |