| Full Scale Ranges |
± 20 nA to ± 20 mA in decade steps, 1 pA maximum resolution |
| Maximum Input Current Without Damage |
± 25 mA |
| Measurement Uncertainty |
± 0.05 % of Reading ± 2 Least Significant Digits |
| A/W Setting |
0.100 to 1.000 A/W in increments of .005 A/W |
| Input Impedance (DC to 2 KHz) |
Zero Ohms Virtual Ground, Single Ended |
| Input Capacitance |
25 pF |
| Output Impedance |
100 Ohms |
| Bias Voltage |
Selectable from - 14 V to + 14 V in 6.5 mv increments |
| Analog Output Port |
± 2 V corresponds to ± 20 000 counts of range in use |
| Noise and Drift |
< ± 1 pA/5 seconds on most sensitive range |
| Background Cancelation |
± 200 % of the range in use |
| Analog Output Port Frequency Response |
DC to 2 KHz, most sensitive range, DC to 40 KHz, least sensitive range |
| Rs-232 Interface |
9600 Baud, N-8-1, 3 wire, Bi-directional, Cable Provided |
| Display |
4 1/2 Digit LCD, 0.4 “ high |
| Power Requirements |
Rechargeable Ni-mH batteries provide approximately 10 hours of use |
| External Power Supply/Charger |
85 - 250 VAC, 50-60 Hz, < 9 VA |
| Mains Adaptors |
Adaptors provided for US, Continental Europe, Great Britain and Australia |
| Operating Temperature Range |
0 - 40 C |
| Dimensions |
5.5” W x 2.5” H x 8.5” L (140 x 63 x 215 mm) |
| Weight |
2 Lbs., 0.9 kg (excluding external power supply) |
| Interconnecting cable supplied |
Rs-232, 14 feet max length |
| CE Certification |
Yes |
| Accessories Provided |
Rs-232 cable, Power Supply/Charger, Operating Manual |
| Standard Warranty |
Two years, Components and Workmanship, 30 Day Satisfaction Guarantee |