| TCP312 w/TCPA300 | TCP305 w/TCPA300 | TCP303 w/TCPA300 | TCP404XL w/TCPA400 |
Bandwidth (-3 dB) | DC – 100 MHz | DC – 50 MHz | DC – 15 MHz | DC – 2 MHz *1 |
Risetime | =3.5 ns | =7 ns | =23 ns | =175 ns |
Maximum Current Ratings: |
High Current Sensitivity Range DC (continuous) RMS (sinusoidal) Peak Pulse | 10 A/V Range 30 A 21.2 A 50 A | 10 A/V Range 50 A 35.4 A 50 A | 50 A/V Range 150 A 150 A 500 A | 1 A/mV Range 500 A (750 A *2) 500 A 750 A |
Low Current Sensitivity Range DC (continuous) RMS (sinusoidal) Peak Pulse | 1 A/V Range 5 A 3.5 A 50 A | 5 A/V Range 25 A 17.7 A 50 A | 5 A/V Range 25 A 17.7 A 500 A | N/A |
DC Accuracy (Operating temp 0 °C to 50 °C) | ±3% of reading | ±3% of reading | ±3% of reading (10 °C to 50 °C) +3%/-6% of reading (0 °C to <10 °C) | ±3% of reading |
DC Accuracy, Typical (Operating temp 23 °C ±5 °C) | ±1% of reading | ±1% of reading | ±1% of reading | ±1% of reading |
Nominal |
Maximum Bare Wire Voltage | For Use With Insulated Wires Only | 600 VRMS CAT I & II 300 VRMS CAT III |
Lowest Measurable Current (at ±3% accuracy at DC) | 1 mA | 5 mA | 5 mA | 1 A |
Scope Set To 1 mV/div and 20 MHz BW Limited | Scope Set To 1 mV/div and 20 MHz BW Limited |
Insertion Impedance (See Curves Below) | 0.08Ω at 1 MHz 0.15Ω at 10 MHz 0.27Ω at 50 MHz 0.7Ω at 100 MHz | 0.035Ω at 1 MHz 0.12Ω at 10 MHz 0.4Ω at 50 MHz | 0.01Ω at 1 MHz 0.025Ω at 5 MHz 0.1Ω at 15 MHz | 0.1ΩO at 10 kHz 0.6Ω at 100 kHz 8Ω at 1 MHz 16Ω at 2 MHz |
Typical |
Maximum Amp-Second Product (Based on Amplifier Range setting) | 50 A*µS – 1 A/V 500 A*µS – 10 A/V | 500 A*µS – 5 A/V NA – 10 A/V | 3,000 A*µS – 5 A/V 15,000 A*µS – 50 A/V | NA – 1 A/mV |
AC Coupling Low Frequency Bandwidth | <7 Hz (Low Pass - 3 dB point) |
Displayed RMS Noise (at 20 MHz Bandwidth Limit) | =250 µA RMS | =1.25 mA RMS | =2.5 mA RMS | =250 mA RMS |
Signal Delay (Delay to Output BNC) | 17 ns | 19 ns | 53 ns | 103 ns |
Inputs (probe amplifier) | 1 |
Probe Open Indicator | YES |
Overload Indicator | YES |
Termination Indicator | YES |
Non-Compatible Probe Indicator | YES |
Safety Certifications |
U.S. NRTL Listing | UL3111-1 (Amplifier) | UL3111-2-032 ; UL3111-1 (Probe and Amplifier) |
Canadian Certification | CAN/CSA C22.2 No.1010.1 (Amplifier) | CAN/CSA C22.2 No.1010.2.032 CAN/CSA C22.2 No.1010.1 (Probe and Amplifier) |
European Union Compliance | EN61010-1/A2 (Amplifier) | EN61010-1/A2; EN61010-2-032 EN61010-1/A2 (Probe and Amplifier) |
Other | IEC61010-1/A2 (Amplifier) | IEC61010-2-032 IEC61010-1/A2 (Probe and Amplifier) |
Electromagnetic Compatibility | Meets EC Council Directive 89/336/EEC, FCC Part 15, Subpart B Class A, AS/NZS 2064.1/2. |
Power Requirements (TCPA300/TCPA400 Amplifiers) | 90 V to 264 V; 47 to 440 Hz; 50 W Maximum CAT II (Auto Switch) |
Power Requirements (Probes) | Requires TCPA300 Amplifier | Requires TCPA400 Amplifier |
Probe Model | TCP312 | TCP305 | TCP303 | TCP404XL |
Warranty | 1 Year |
Probe Mechanical Characteristics |
Probe Cable Length | 1.5 meters (60 inches) | 2 meters (78.7 inches) | 8 meters (315 inches) |
Probe Jaw Size (Max Conductor Size) | 3.8 mm (0.15 inches) | 21 mm x 25 mm (0.83 x 1.0 inches) |
Length: Width: Height: Weight: | 20 cm (7.87 inches) 1.6 cm (0.625 inches) 3.2 cm (1.25 inches) 0.15 kg (0.33 lb.) | 26.8 cm (10.55 inches) 4.1 cm (1.60 inches) 15.6 cm (6.13 inches) 0.66 kg (1.45 lb.) | 26.8 cm (10.55 inches) 4.1 cm (1.60 inches) 15.6 cm (6.13 inches) 0.88 kg (1.90 lb.) |
TCPA300 and TCPA400 Mechanical Characteristics |
Length: Width: Height: Weight: | 17.3 cm (6.8 in) 16.7 cm (6.6 in) 9.14 cm (3.6 in) 1.14 kg (2.5 lb.) |
Environmental Characteristics |
Temperature - Operating | 0 °C to +50 °C (32 °F to 122 °F) |
Temperature - Nonoperating | -40 °C to +75 °C (-40 °F to 167 °F) |
Humidity – Operating | 5% to 95% R.H.to +30 °C (86 °F) 5% to 85% R.H.+30 °C to +50 °C (86 °F to 122 °F) |
Humidity – Nonoperating | 5% to 95% R.H.to +30 °C (86 °F) 5% to 85% R.H.+30 °C to +75 °C (86 °F to 167 °F) |
Altitude – Operating | 2000 m (6800 ft) maximum |
Altitude – Nonoperating | 12,192 m (40,000 ft) maximum |