Dri-Block Temperature Calibrators
The Tecal Series of Dri-Block Calibrators are used as portable temperature calibrators in a wide variety of industries for checking the calibration of thermocouples, RTDs and liquid fi lled sensors. The checking of temperature is vital in numerous industrial processes – engines, machines, boilers, pumps, storage rooms, air conditioners, compressors used in power stations, chemical plants, refi neries, off shore platforms, ships, steel works and instrument companies are all typical application areas.
Tecal H Range Dri-Block Calibrator
This range of temperature calibrators is ideal for advanced temperature calibration work where programmability, switch testing and ramp/dwell functionality from the unit’s display and user interface are required.
Each Tecal H model is shipped with one insert, soft-sided carrying case, calibration certificate and TechneWorks software with cable.
H Version: Set required Temperature and Calibrate; also programmable for switch test and ramp rate.
Tecal 140H
Using advanced peltier technology and state-of-the-art mechanical and electrical components, the Tecal 140H can rapidly cover the complete temperature range where speed is of the essence without loss of accuracy or stability. The heater block, with a defi ned measuring zone of 5 cm from the bottom of the block, allows for diff erent length/active area sensors. Although the quoted accuracy and stability is ±0 .3°C & ±0.05°C respectively, one can often achieve an accuracy of ±0.1°C and a stability of ±0.02°C with high class RTDs and thermocouples.
TechneWorks Software
TechneWorks is an easy to use control software for many of our RS-232 compatible products. This software has been designed to supersede the existing Calsoft and Thermsoft software packages.
It will allow customers to carry out all the tasks previously available on Calsoft and Thermsoft, plus many new options on one new package. It is fully compatible with all Techne RS-232 Calibration products including all current models and older discontinued models such as the TU-20C.
TechneWorks software allows connection to a range of Precision Thermometers, such as the Techne Accu-Temp, ASL model F250, Accu-Temp II and the Cropico 3000 series. TechneWorks software enables you to calibrate thermometers and sensors by comparison to one of the above listed thermometers using a Techne liquid bath or dry block as a temperature source. Alternatively, the bath or dry block calibrator display can be used as the reference. After data is collected, a calibration certificate can be printed.
Software Features:
- Create, open and save programs with up to 20 set points.
- Specify either °C or °F.
- Specify ramp rates and hold times.
- Log data from the instrument while connected to the computer and export the data in an Excel spreadsheet.
- Open, save, view and print logged data.
- Perform a calibration where temperature of the probe is manually entered.
- Perform an automatic calibration routine where the temperature reference probes data is added automatically.
- Run a program in real time mode.
- Specify the logging interval from every 5 seconds to 60 seconds
- Free download from website
- Perform a calibration using the User Probe Interface
- Perform a calibration requiring a switch test
- Send programs to the calibrator
- Retrieve programs from the calibrator
- Retrieve test results from the calibrator
- Erase test results (downloadable free of charge from www.techneusa.com)
- Note: CD with software included with Tecal H and TU-20HT models