SPECIFICATIONS |
Specifications apply after 30 minute warm-up, ambient 5°C to 40°C |
Frequency Range: |
10MHz to 6000MHz |
Setting Resolution: |
10Hz by direct keyboard entry, or in user-set increments of 10Hz to 999 999MHz by rotary control or increment decrement keys. |
Display Resolution: |
10Hz |
Phase Noise: |
500MHz Carrier: <-110dBc/Hz (typ) @ 20kHz offset; <-120dBc/Hz (typ) @ 100kHz
3GHz Carrier: <-95dBc/Hz (typ) @ 20kHz offset; <-110dBc/Hz (typ) @ 100kHz offset
6GHz Carrier: <-89dBc/Hz (typ) @ 20kHz offset; <-104dBc/Hz (typ) @ 100kHz offset |
Residual FM: |
Equivalent peak deviation for 300Hz to 3?4kHz B/W: 12Hz at 500MHz carrier |
Settling Time: |
<8ms to settle within 100Hz or 0.1ppm of final frequency, if greater. |
Options: |
Internal or External (via rear panel BNC) |
Internal Accuracy: |
+/- 1 ppm over temperature range 1 5°C to 30°C; +/- 2 ppm over 5°C to 40°C. |
Internal Stability: |
< +/-1ppm/year ageing. |
Internal Ref. Out: |
10MHz from 50 Ohms, amplitude 2V pk-pk into 50 Ohms. |
External Ref In: |
10MHz into 50 Ohms, amplitude 2V pk-pk to 5V pk-pk. |
Output Level Range: |
-110dBm to +7dBm (01V to 500mV into 50 W). |
Setting Resolution: |
0.1dB (or 0.01µV to 1mV) by direct keyboard entry, or in user-set increments of 01dB to 100dB (or 0.01mV to 100mV) by rotary control or increment decrement keys. |
Accuracy: |
Better than 2dBm. |
Harmonics: |
<-25dBc at +7dBm. <-30dBc@levels <=0dBm 30 to 6000MHz,
<-25dBc@levels <=0dBm 10 to 30MHz |
Sub-Harmonics: |
<3000MHz : None. >3000MHz : |
Non-Harmonic Spurii: |
10kHz offset 10 to 3000MHz (Note 1)
<-44dBc >10kHz offset 3000MHz to 6000MHz (Note 2).
Note 1 - <-45dBc >10kHz offset 1900 2150MHz
Note 2 - <-39dBc >10kHz offset 3800 4300MHz. |
Carrier Leakage: |
<0.5uV generated into a 50 W load by a 2 turn 25mm loop, at 25mm from the generator with output set to <-10dBm into a 50 W sealed load. |
Output Impedance: |
50 W |
Output Connector: |
TYPE N |
Reverse Protection: |
50V DC |
Output Switch: |
RF OUT on-off switch with LED for ON status. |
Step Sweep |
Step frequency and/or amplitude according to a formula over a specified number of points. Formula Specifies: Frequency start/stop. Amplitude start/stop.
Dwell time at each step programmable 0.01 to 10.000sec |
Max Points: |
1000 |
Sweep Run: |
Continuous or single. Sweep up or down |
Step Spacing: |
Linear or logarithmic |
Sweep Triggering: |
Manual, Ext. signal, timed (0.1 999.9sec) or via remote interface |
Sync Signal: |
(Output Stable) available during dwell time. Programmable to be high or low |
List Sweep |
As for Step Sweep except that a user defined table of frequency, amplitude and dwell time values defines the steps. The table can be created within the instrument or downloaded via the remote interfaces. |
Max Points: |
1000 |
List Storage: |
Up to 16 Sweep Lists can be stored permanently within the instrument |
Point Trigger |
Each point in a sweep (step or list) can be subject to a trigger event rather than a dwell time. |
Point Triggering |
Manual, Ext. signal or via remote interface. |
|
A table of frequency/gain pairs allows the user to modify the generator output level with respect to frequency to calibrate an entire test set up or improve the calibration of the generator alone. The table can be created within the instrument or downloaded via the remote interfaces. |
Max Points: |
100 |
Trig In |
DC coupled External Trigger Input signal used for step Sweep changes in Point Trigger mode |
Input Threshold: |
1.65V nominal. Trigger polarity can be set to Negative Edge or Postive Edge |
Input Protection |
Maximum/minimum external applied voltage is +6V or 1V. |
Rear panel output SYNC signal goes to its active state when generator output frequency & level have settled within specification after a step change during Sweep. SYNC returns to inactive state at end of specified dwell period |
Active Output Level: |
+5V (Active state set to Pos ) or 0V (Active state set to Neg ) |
Output Impedance: |
50 Ohm Minimum load impedance is also 50 Ohm |
Output Protection: |
Output will withstand accidental short circuit to ground and applied external voltages up to +5V. |
Full digital remote control facilities are available through the RS232, USB, LAN and GPIB interfaces |
9-pin D connector. Variable Baud rate can be set between 1200 and 115200 Baud maximum. |
USB 2.0 connection (backwards compatible with USB 1.x). Operates as a virtual COM port |
Conforming with IEEE4881 and IEEE4882. |
Standard 10/100 base-T hardware connection. ICMP and TCP/IP Protocol for connection to Local Area Network or direct connection to a single PC |
Display: |
20 character x 4 row backlit alphanumeric LCD |
Data Entry: |
Keyboard selection of all major parameters. Value entry by character scroll using rotary control or up/down keys, or value stepping in user-selected increment values using rotary control or up/down keys |
Stored Settings: |
Up to 12 complete instrument set-ups plus 16 sweep lists may be stored in non volatile memory. |
Size: |
2U (86.5mm) height; half-rack (213mm) width; 350mm long. |
Weight: |
3.65 kg. (8 lb) |
Power: |
110-240VAC 10% 50/60Hz; 100-120VAC 10% 400Hz; 60VA max.
Installation Category II |
Location |
Built-in tilt feet for bench-top use. Rack mountable with optional mount |
Options: |
19 inch rack mounting kit. |
Operating Range: |
+5°C to 40°C, 20-80% RH. |
Storage Range: |
20°C to + 60°C. |
Environmental: |
Indoor use < 2000m, Pollution Degree 1. |
Safety: |
Complies with EN61010-1. |
EMC: |
Complies with EN61326 |