FREQUENCY |
Frequency Range |
0.005Hz to 5MHz in 7 overlapping decade
ranges with fine adjustment by verniers. |
Vernier Range |
1000:1 on each range. |
FREQUENCY LOCKING |
Operating Range |
0.5Hz to 5MHz with vernier setting within
10% to 100% of range maximum. |
Locking Accuracy |
Better than 0.01% of displayed value. |
DISPLAY |
LCD, 8 digits (11mm high) plus 14 annunciators |
METER FUNCTIONS (generator mode) |
Frequency |
Auto-ranging reciprocal measurement
giving4 digit resolution for frequencies
down to 1Hz; maximum resolution is
0.001Hz. Accuracy ±1 digit 0.2Hz to
5MHz; accuracy ±1% of range full scale
below 0.2Hz |
Amplitude |
Display shows peak-to-peak amplitude
or rmsvalue. Display corrected for attenuator
setting. 3-digit resolution, accuracy typically
5% of range |
DC Offset |
3-digit resolution; accuracy typically ±2% of
setting ±1 digit. Display corrected for attenuator setting |
OUTPUTS |
2mV to 20V peak-peak open circuit (1mV to 10V peak-
peak into 50 ) in four switch-selectable ranges with 20dB
vernier control within each range |
Attenuator Ranges |
0dB, -20dB, -40dB, -60dB |
DC Offset Range |
±10V from 50 . DC offset plus signal peak limited to ±10V
(±5V into 50 ). DC offset plus waveform attenuated
proportionally by the attenuator. |
Alternative output socket offering the same facilities as
the 50 socket. |
0 to 5V TTL/CMOS logic levels capable of driving 2
standard TTL loads. Frequency, symmetry and phase
as main outputs |
3V ramp from 600 |
EXTERNAL COUNTER |
Frequency Range |
5Hz to 20MHz, fully autoranging |
Input Sensitivity |
50mV rms (sinewave) |
Input Impedance |
1M /25pF |
Measurement Time |
Selectable 0.5s or 5s |
Resolution |
6 digits in 0.5s; 7 digits in 5s |
Accuracy |
±1 digit ± timebase accuracy |
Timebase Accuracy |
±10ppm initial error; ±5ppm/year ageing rate; typically
less then 0.5ppm/ o C. Adjustment point for closed-case
recalibration |
OPERATING MODES |
Specifications apply for the top decade of each frequency range and maximum
output into 50 termination. |
Distortion |
<0.5% on 500, 5k and 50k ranges; <1% on 5, 50 and
500k ranges; all harmonics >25dB below fundamental
on 5M range. |
Amplitude Flatness |
±0.2dB to 200kHz; ±2dB to 5MHz. |
Linearity |
Better than 99% to 200kHz |
Rise/Fall Times |
<45ns |
Mark - Space Ratio |
1:1 ± 1% to 100kHz |
Range |
±10V unterminated |
Symmetry Range |
Variable typically 1:9 to 9:1 (on top decade of each
range), frequency divided by 10. |
SWEEP MODES |
Sweep Range |
>1000:1 within each range. |
Sweep Rate |
Adjustable, typically 20ms to 20 secs. |
Sweep Mode |
Linear or logarithmic. Sweep start and stop frequencies
displayed by button press. |
Input Impedance |
10k |
Input Sensitivity |
0 to 3V for 1000:1 sweep |
Max. Input Voltage |
±10V |
Sweep Linearity |
Better than 1% |
Max. Voltage Rate |
0.1V/us |
AMPLITUDE MODULATION |
Depth |
Variable 0 to 100% |
Frequency |
400Hz (internal). DC to 20kHz (external). |
External Sensitivity |
Approx. 2V peak-to-peak for 50% modulation. |
GENERAL |
AC Input |
220 to 240 volts ±10% or 110 to 120 volts ±10%
50/60Hz, Installation Category II. |
Consumption |
25VA max. |
Operating Range |
+5 o C to +40 o C, 20% to 80% RH. |
Storage Range |
-10 o C to +65 o C |
Environmental |
Indoor use at <2000m, Pollution degree 2 |
Safety |
Complies with EN601010-1 |
EMC |
Complies with EN61326 |
Size |
260(W) x 88(H) x 235(D)mm , excluding handle and feet. |
Weight |
1.9kg approx. |