TTi TG1000 Specifications |
Specifications apply at 18°- 28°C after 1 hour warm-up, at maximum output into 50Ω . |
FREQUENCY |
All waveforms are derived from a crystal clock using Direct Digital Synthesis. |
Frequency Range |
1mHz to 10MHz (TG1000) or 20MHz (TG2000) |
Resolution |
6 digits or 1mHz |
Accuracy |
±10ppm for 1 year, 18°C to 28°C; ±1mHz below 0.2Hz |
Tempco |
Typically <1ppm/°C outside of 18°C to 28°C |
WAVEFORMS |
Sinewave |
Range |
1mHz to 10MHz/20MHz |
Resolution |
6 digits or 1mHz |
Distortion |
<0.3% THD to 20kHz (typically 0.1%), <-45dBc to
300kHz, <-35dBc to 20MHz (typically <-40dBc) |
Spurii |
Non harmonically related spurii <-55dBc to 1MHz,
<(-55dBc + 6dB/octave) 1MHz to 20MHz |
Output Level |
5mV to 20V pk-pk from 50 Ω or 600Ω |
Squarewave |
Range |
1mHz to 10MHz/20MHz |
Resolution |
6 digits or 1mHz |
Symmetry |
variable 20% to 80% in 1% steps |
Aberrations |
<5% + 2mV |
Rise & Fall Times |
<22ns |
Output Level |
5mV to 20V pk-pk from 50Ω or 600Ω |
Triangle |
Range |
1mHz to 1MHz |
Resolution |
6 digits or 1mHz |
Linearity error |
<0.5% to 100kHz |
Output Level |
5mV to 20V pk-pk from 50Ω or 600Ω |
Positive and Negative Pulse |
Range |
1mHz to 10MHz/20MHz |
Resolution |
6 digits or 1mHz |
Symmetry |
variable 20% to 80% in 1% steps |
Aberrations |
<5% + 2mV |
Rise & Fall Times |
<22ns |
Output Level |
2.5mV to 10V pk-pk from 50/600Ω positive or negative
only pulses with respect to the DC Offset baseline |
MODULATION MODES |
Continuous |
Continuous cycles of the selected waveform are output at the selected frequency. |
Gated |
Non phase-coherent signal keying - output is On while Gate signal is high and Off
while low. |
Carrier frequency |
From 0.1Hz to 10MHz/20MHz |
Carrier waveforms |
All |
Trigger rep. rate |
dc to 100kHz external, dc to 5kHz internal |
Gate source |
Front panel MAN TRIG key, Internal Gate Generator,
TRIG/GATE input, or Remote Interface |
Sweep |
Carrier waveforms |
All |
Sweep Mode |
Linear or logarithmic, single or continuous |
Sweep Width |
0.2Hz to 10MHz/20MHz in one range. Phase continuous.
Independent setting of the start and stop frequency |
Sweep Time |
50ms to 999s (3 digit resolution) |
Markers |
Available from AUX output. Variable during sweep |
Sweep Trigger
source |
The sweep may be free run or triggered from: front panel
MAN TRIG key, TRIG/GATE input, or Remote Interface |
Amplitude Modulation |
Carrier frequency |
1mHz to 10MHz/20MHz |
Carrier waveforms |
All |
Modulation source |
VCA IN socket |
Frequency Shift Keying (FSK) |
Phase coherent switching between two selected frequencies at a rate defined by the
switching signal source. |
Carrier frequency |
1Hz to 10MHz/20MHz |
Carrier waveforms |
All |
Switch rate |
dc to 5kHz (internal), dc to 1MHz (external) |
Switching signal
source |
Front panel MAN TRIG key, Internal Trigger Generator,
TRIG/GATE input, or Remote Interface |
Tone |
The tone is output while the trigger signal is high, and stopped when the trigger signal
is low. The next tone is output when the trigger signal goes high again. |
Carrier waveforms |
All |
Frequency list |
Up to 16 frequencies between 1Hz and 10MHz/20MHz |
Min. switching time |
1ms per tone |
Switching source |
Front panel MAN TRIG key, Internal Trigger Generator,
TRIG/GATE input, or Remote Interface |
Internal Trigger/Gate Generator |
Period |
0.2ms to 999s (resolution 0.2 ms) |
Waveform |
Square wave (1:1 duty cycle) |
MAIN OUTPUT |
Output Impedance |
50Ω or 600Ω switchable |
Amplitude |
5mV to 20V pk-pk open circuit (2.5mV to10V into 50/600Ω )
Output can be specified as V-HiZ (open circuit value) or V (potential difference) in
pk-pk, RMS or dBm. Note that in positive or negative Pulse modes the amplitude range
is 2.5mV to 10V pk-pk O/C. |
Accuracy |
±3% ±1mV at 1kHz into 50Ω /600Ω |
Flatness |
±0.2dB to 500kHz; ±1dB to 10MHz; ±2dB to 20MHz |
DC Offset |
±10V from 50Ω /600Ω . DC offset plus signal peak limited to
±10V. Accuracy ±3% ±10mV |
Resolution |
3 digits for both amplitude and offset |
AUXILIARY OUTPUT |
Multi-function output user definable to be any of the following: |
Waveform Sync |
Outputs a 50% duty cycle squarewave at the main wave-
form frequency |
Trigger Out |
Outputs a replica of the current trigger signal |
Sweep Sync |
Outputs a trigger signal at the start of sweep (for synchro-
nising an oscilloscope or chart recorder). Can additionally
output a sweep marker. |
Signal Levels |
Output Impedance 50Ω nominal. Logic levels of <0.8V and
>3V. Sweep Sync is a 3 level waveform, low at start of
sweep, high at end of sweep, with a narrow 1V pulse at the
marker point |
INPUTS |
Ext Trig/Gate |
Frequency Range |
DC to 1MHz for FSK; DC to 100kHz for Gate; DC to 2.5kHz
for Tone and Sweep |
Signal Range |
Nominal TTL level threshold; maximum input ±10V |
Min. Pulse Width |
100ns for Gate/FSK; 0.2ms for Sweep and Tone |
Input Impedance |
Typically 10kΩ |
VCA In |
Frequency Range |
DC - 100kHz |
Signal Range |
2.5V for 100% level change at maximum output |
Input Impedance |
Typically 6kΩ |
INTERFACES (TG2000 only) |
Full remote control facilities are available through the RS232 or USB interfaces. |
RS232 |
Variable Baud rate (19200 max), 9-pin D-connector. As
well as operating in a conventional RS-232 mode the inter-
face can be operated in addressable mode whereby up to
32 instruments can be addressed from one RS-232 port |
USB |
Standard USB hardware connection. Conforming USB 1.1 |
GENERAL |
Display |
20 character x 4 row alphanumeric LCD |
Data Entry |
Keyboard selection of mode, waveform etc.; value entry di-
rect by numeric keys or by rotary control. |
Stored Settings |
Up to 9 complete instrument set-ups may be stored in
battery-backed memory. |
Size & Weight |
260(W) x 88(H) x 235(D) mm; 2kg (4.5lb) |
Power |
100V, 110-120V or 220-240V ±10% 50/60Hz, adjustable internally. 40VA max. Installation Category II. |
Operating Range |
+5°C to 40°C, 20-80% RH |
Storage Range |
-20°C to +60°C |
Environmental |
Indoor use at altitudes up to 2000m, Pollution Degree 2 |
Safety & EMC |
Complies with EN6010-1 and EN61326 |