| Velocity |
| Range (AVM410) |
0 to 4,000 ft/min (0 to 20 m/s) |
| Range (AVM430, AVM440) |
0 to 6,000 ft/min (0 to 30 m/s) |
| Accuracy (AVM410) |
±5% of reading or ±5 ft/min (±0.025 m/s), whichever is greater |
| Accuracy (AVM430, AVM440) |
±3% of reading or ±3 ft/min (±0.015 m/s), whichever is greater |
| Resolution |
1 ft/min (0.01 m/s) |
| Duct Size (AVM430, AVM440) |
| Dimensions |
1 to 250 inches in increments of 0.1 in.
(1 to 635 cm in increments of 0.1 cm) |
| Volumetric Flow Rate (AVM430, AVM440) |
| Range |
Actual range is a function of velocity, and duct size |
| Temperature |
| Range (AVM410, AVM430) |
0 to 200°F (-18 to 93°C) |
| Range (AVM440) |
14 to 140°F (-10 to 60°C) |
| Accuracy |
±0.5°F (±0.3°C) |
| Resolution |
0.1°F (0.1°C) |
| Relative Humidity (AVM440 only) |
| Range |
0 to 95% RH |
| Accuracy |
±3% RH |
| Resolution |
0.1% RH |
| Wet Bulb Temperature (AVM440 only) |
| Range |
40 to 140°F (5 to 60°C) |
| Resolution |
0.1°F (0.1°C) |
| Dew Point (AVM440 only) |
| Range |
5 to 120°F (-15 to 49°C) |
| Resolution |
0.1°F (0.1°C) |
| Instrument Temperature Range |
| Operating (Electronics) |
40 to 113°F (5 to 45°C) |
| Model AVM410, AVM430 Operating (Probe) |
0 to 200°F (-18 to 93°C) |
| Model AVM440 Operating (Probe) |
14 to 140°F (-10 to 60°C) |
| Storage |
-4 to 140°F (-20 to 60°C) |
| Data Storage Capabilities (AVM430, AVM440) |
| Range |
12,700+ samples and 100 test IDs |
| Logging Interval (AVM430, AVM440) |
| 1 second to 1 hour |
| Time Constant (AVM430, AVM440) |
| User selectable |
| External Meter Dimensions |
| 3.3 in. x 7.0 in. x 1.8 in. (8.4 cm x 17.8 cm x 4.4 cm) |
| Meter Weight with Batteries |
| 0.6 lbs. (0.27 kg) |
| Meter Probe Dimensions |
| Probe Length |
40 in. (101.6 cm) |
| Probe Diameter of Tip |
0.28 in. (7.0 mm) |
| Probe Diameter of Base |
0.51 in. (13.0 mm) |
| Articulating Probe Dimensions |
| Articulating Section Length |
7.8 in. (19.7 cm) |
| Diameter of Articulating Knuckle |
0.38 in. (9.5 mm) |
| Power Requirements |
| Four AA-size batteries or AC adapter |