TPI 343C1 Dual Input K-Type Thermocouple Thermometer with Field Calibration with two GK11M, A304, A340 soft pouch (For High Temp. over 400 F)
TPI contact temperature testers are accurate, reliable, and built to last. Model 343 can be auto field calibrated to system accuracy (tester plus probe) of +/- 1°F in less than 10 seconds in ice water.
AUTO FIELD CALIBRATION Perform ice bath calibration to achieve ±1°F system accuracy within the 30°F to 120°F temperature range. Calibration is an easy two-step process performed with instrument keypad, no additional tools needed.
Field calibration for thermometers TPI 343 FIELD CALIBRATION
The 343 has the ability to be field calibrated. Accuracy of +/-1°F within the 30°F to 120°F range can be achieved by performing an ice bath calibration.
The temperature of an ice bath is approximately 32°F (0°C). Fill a plastic or metal container with crushed ice and add clean water to a depth of at least 4 inches. Stir the ice and water for 2 to 3 minutes prior to performing calibration to ensure the water is completely chilled. Make certain there is plenty of ice in the mixture and always use clean water. Distilled water works well.
1. Step-by-Step Procedures
- Connect both temperature probes to the 343.
- Press and hold down the MIN/MAX and HOLD buttons and turn the 343 on. CALF will be displayed and the 343 will enter calibration mode and begin cycling between T1 and T2.
- Insert the probes into the ice bath and allow the readings to stabilize.
Once the reading on both inputs has stabilized, press the HOLD button. SAVE will be displayed and calibration is complete.