INPUT |
Voltage
|
264 |
100 - 120 VAC / 200 - 240 VAC ± 10% Auto Range |
266 |
100 - 240 VAC ± 10% Full Range |
Frequency |
50/60 Hz ± 5% |
Fuse
|
264 |
10A / 250 VAC Slow-Blow |
266 |
15A / 250 VAC Slow-Blow |
GROUND BOND TEST MODE |
Output Rating
|
264 |
3.0 - 40.0 AAC |
266 |
3.0 - 60.0 AAC |
Resolution: 0.1 A |
Regulation: ± (2% of setting + 0.1A) |
264 |
Voltage 6 VAC (fixed) |
266 |
Voltage 12 VAC (fixed) |
Output Frequency |
50/60 Hz user selectable
Accuracy: ± 0.1% |
Resistance Limit Settings
|
264 |
0 - 150 mΩ for 30.1 - 40.0 A
0 - 200 mΩ for 10.1 - 30.0 A
0 - 600 mΩ for 3.0 - 10.0 A |
266 |
0 - 150 mΩ for 30.1 - 60.0 A
0 - 200 mΩ for 15.1 - 30.0 A
0 - 600 mΩ for 3.0 - 15.0 A |
Resolution: 1 mΩ
Accuracy: ± (2% of setting + 2 mΩ) |
Offset Limit Settings
|
0 - 100 mΩ
Resolution: 1 mΩ
Accuracy: ± (2% of setting + 2 mΩ) |
Dwell Timer |
0, 0.5 - 240.0 sec, (0=continuous), 0.1 sec/step |
Ramp Timer |
0.1 sec fixed |
Measurement Current
|
264 |
0.0 - 40.0 AAC |
266 |
0.0 - 60.0 AAC |
Resolution: 0.1 A
Accuracy: ± (3% of reading + 0.1 A) |
Ohmmeter
|
264 |
0 - 600 mΩ |
|
Resolution: 1 mΩ
Accuracy: ± (3% of reading + 3 mΩ) for 3 - 5.9 A |
266 |
0 - 600 mΩ |
|
Resolution: 1 mΩ
Accuracy: ± (3% of reading + 3 mΩ) for 3 - 5.9 A
± (2% of reading + 2 mΩ) for 6 - 40 A |
GENERAL SPECIFICATIONS |
Memories |
5 |
Remote I/O
|
Input: |
Test, Reset, Interlock |
Output: |
Pass, Fail, Test-in-Process |
Hardware Interlock - a relay on the high voltage output opens when the Interlock signal is disabled. |
Voltage Drop Display (optional) |
Display the voltage drop across the circuit instead of the resistance measurement. |
Voltage Limit Settings
|
264 |
0.00 - 6.00 VAC |
266 |
0.00 - 9.00 VAC |
Resolution: 0.01 V
Accuracy: ± (2% of setting + 0.02 V) |
Offset Limit Settings
|
264 |
0.00 - 4.00 VAC |
266 |
0.00 - 6.00 VAC |
Resolution: 0.01 V
Accuracy: ± (2% of setting + 0.02 V) |
Security
|
Option to turn On or Off, when On you can switch between two security levels:
1. Run - Operator can only run a test. No ability to change memory locations or edit test parameters.
2. Mem - Operator can run a test and change memory locations. No ability to edit test parameters. |
Safety Mark |
CE/cTUVus |
Dimensions
(W x H x D) |
264 |
8.5" x 3.5" x 11.81" (215 x 88 x 300 mm) |
266 |
16.93" x 5.20" x 11.81" (430 x 132 x 300 mm) |
Weight
|
264 |
9.25 lbs. (4.2 Kg) |
266 |
20.25 lbs. (9.2 Kg) |