Specification for F45 Frequency Meter |
Display |
Type |
7-segment, red LED |
Height |
0.56" (14.2mm) |
Decimal Point |
Jumper-selectable 2-position (corresponding to resolution desired) |
Overrange indication |
Most significant digit = "1"; other digit blank |
Polarity |
Automatic, with "-" indication,
"+" indication implied |
Power Requirements |
AC Voltages |
120 or 220VAC, ±10% 50/60Hz |
DC Voltages |
9-32 DCV, ±1% |
Power Consumption |
3VA |
Environmental |
Operating Temperature |
0 to 55°C |
Storage Temperature |
-10 to 60°C |
Relative Humidity |
0 to 85% non-condensing |
Warm-up Time |
Less than 20 minutes |
Inputs |
Thermocouple |
J, K, E, T, R, and S |
RTD |
Platinum 100 (.00385 alpha), 2, 3 or 4 wire |
Millivolt |
±84mV reading of uncompensated mV |
Input Impedance |
7M:(typical) |
Conversion Rate |
2-1/2 times per second |
Open Thermocouple Detection |
-1 on display, -40nA bias on thermocouple |
General Specifications |
Display |
Type |
7-segment, red LED |
Height |
0.56" (14.2mm) |
Decimal Point |
user-programmable,
internally or on the terminal block |
Overrange indication |
most significant digit = "1"; other digits
blank |
Polarity |
Automatic, with "-" indication,
"+" indication implied |
Power Requirements |
AC Voltages |
120 or 220VAC, ±10% 50/60Hz |
DC Voltages |
9-32DCV, ±1% |
Power Consumption |
F35: 3VA, F45: 2VA |
ISOLATION |
250V RMS MAX |
Accuracy @25°C |
F45 DC Process/Voltage |
±0.02% of reading ± 1 count |
F35 DC Process/Voltage |
±0.1% of reading ± 1 count |
F45 DC Current |
±0.05% of reading ±1 count
2A ± 5 counts 5A ± 5 counts |
F35 DC Current |
±0.1% of reading ± 1 count
2A ± 5 counts, 5A ± 5 counts |
F45 AC Voltage/Current |
±0.5% of reading ± 35 counts |
F35 AC Voltage/Current |
±1% of reading ± 5 counts (45Hz-1KHz) |
Environmental |
Operating Temperature |
0 to 55°C |
Storage Temperature |
-10 to 60°C |
Relative Humidity |
0 to 85% non-condensing |
Warm-up Time |
Less than 15 minutes |
Noise Rejection |
NMRR |
F35=50dB, 50/60Hz; F45=60dB, 50/60Hz |
CMRR |
(w/1K unbalanced @ 60Hz) 90dB min. |
A To D Conversion |
Technique |
Dual slope integration |
Rate |
F35=3 samples per second, nominal;
F45=2.5 samples per second,nominal |
Mechanical |
Bezel |
3.78" x 1.89" x .51" |
Depth |
2.94" |
Panel Cutout |
3.62" X 1.77" |
Case Material |
94V-1, UL rated Noryl® |
Weight |
9.0oz (255.1g) |
Temperature Coefficient |
F35 AC/TRMS (Voltage/Current) |
(±0.1% ± 0.5 count)/°C |
F35 DC (Voltage/Current/Process) |
(±0.01% ± 0.05 count)/°C |
F45 AC (Voltage/Current) |
(±0.05% ± 0.5 count)/°C |
F45 DC Current |
(±0.01% ± 0.1 count)/°C |
F45 DC (Voltage/Process) |
(±0.05% ± 0.1 count)/°C |