| Inputs |
| Input Coupling |
AC, DC, GND |
| Input Impedance |
1MO±2% || 16Pf±3Pf, |
| Maximum Input voltage |
400V (DC+AC PK-PK,1MO input impedance, X10),CAT I |
| Ch to Ch Isolation (Both channels in same V/div setting) |
>100:1 at 100MHz: (SDS1202DL) <>100:1 at 50MHz:(SDS1102DL)
>100:1 at 25MHz (SDS1052DL)
>100:1 at 10MHz (SDS1022DL) |
| Probe Attenuator |
1X,10X |
| Probe Attenuator Factors Set |
1X,5X,10X,50X,100X, 500X,1000X |
| Vertical System |
| Vertical Sensitivity |
2mV/div -10V/div(1-2-5 order) |
| Channel Voltage Offset Range |
2mV –200mV: ±1.6V 206mV - 10V: ±40V |
| Vertical Resolution |
8 bit |
| Channels |
2 |
| Analog |
200MHz(SDS1202DL) |
| Bandwidth |
100MHz(SDS1102DL)
50MHz(SDS1052DL)
25MHz(SDS1022DL) |
| Single-shot Bandwidth |
200MHz(SDS1202DL)
100MHz(SDS1102DL)
50MHz(SDS1052DL)
25MHz(SDS1022DL) |
| BW Flatness at BNC input |
DC -10% of rated BW: +/- 1dB
10% - 50% of rated BW: +/- 2dB
50% - 100% of rated BW: + 2dB/-3dB |
| Lower frequency limit (AC -3dB) |
=10Hz(at input BNC) |
| Noise Pk-Pk for 3K record |
=0.6 Div for average of 10 Pk-Pk readings, Fixed gain settings
=0.7 Div for average of 10 Pk-Pk readings, Variable gain settings |
| SFDR including harmonics (measured with FFT) |
>=35dB |
| DC Gain Accuracy |
<±3.0%: 5mv/div to 10V/div in Fixed Gain Ranges
<±4.0%: 2mv/div Variable Gain Ranges |
| DC Measurement Accuracy |
All Gain settings =100mv/div : ± [3%* ( |reading|+|offset| ) +1% *of |offset| +0.2div+2mv] |
| DC Measurement Accuracy |
All Gain settings >100mv/div : ± [3%* ( |reading|+|offset| ) +1%* of |offset| +0.2div+100mv] |
| Rise time |
<1.8ns (SDS1202DL )
<3.5ns(SDS1102DL)
<7.0ns (SDS1052DL )
<14ns (SDS1022DL ) |
| Overshoot, Typical (using 500ps pulse) |
<10% with probe or BNC input w/ 50 Ohm feed thru |
| Ch to Ch Skew (both channels in same V/div setting) |
<1ns: SDS1202DL SDS1102DL
<4ns: SDS1052DL
<10ns: SDS1022DL
(Equivalent to 2 minor divisions in smallest t/div) |
| Math operation |
+, -, *, /, FFT |
| FFT |
Window mode: Hanning, Hamming, Blackman, Rectangular
Sampling points: 1024 |
| Bandwidth limited |
20MHz ± 40% (Note: BW limited below
20MHz when using probe in x1) |
| Horizontal System |
| Real Time Sampling Rate |
Single Channel:500MSa/s,Double Channel: 250MSa/s( When timebase faster than 250ns/div) |
| Equivalent Sampling Rate |
50GSa/s (SDS1022DL:10GSa/s) |
| Measure Display Modes |
MAIN, WINDOW, WINDOW ZOOM, ROLL, X-Y |
| Timebase Accuracy |
±100ppm measured over 1ms interval |
| Horizontal Scan Range |
1/2.5/5/25nS/DIV - 50S/DIV (According to the Bandwidth)
Scan: 100mS/DIV ~50S/DIV (1-2.5-5 sequence) |
| Trigger System |
| Trigger Types |
Edge, Pulse Width, Video, Slope, Alternative |
| Trigger Source |
CH1,CH2,EXT,EXT/5,AC Line |
| Trigger Modes |
Auto, Normal, Single |
| Trigger Coupling |
AC, DC, LF rej, HF rej |
| Trigger Level Range |
CH1,CH2: ±6divisions from center of screen
EXT: ±1.2V
EXT/5: ±6V |
| Trigger Displacement |
Pre-trigger:(Memory depth/(2*sampling)),
Delay Trigger: 271.04DIV |
| Trigger Level Accuracy (typical) applicable for the signal of rising and falling time =20ns |
Internal: ±(0.2 div×V/div)( within±4 divisions from center of screen)
EXT: ±(6% of setting + 40 mV)
EXT/5: ±(6% of setting + 200 mV) |
| Trigger Sensitivity |
For fixed gain ranges
1 Divisions: DC-10MHz
1.5 Divisions: 10MHz - Max BW
EXT: 200mVpp DC-10MHz, 300mVpp 10MHz - Max BW
EXT/5: 1Vpp DC-10MHz, 1.5Vpp 10MHz - Max BW |
| Pulse Width Trigger |
Trigger Modes: (>,<, =)positive Pulse Width, (>, <, =)Negative Pulse Width
Pulse Width Range: 20ns – 10s |
| Video Trigger |
Support signal Formats: PAL/SECAM, NTSC
Trigger condition : odd field, even field, all lines, line Num |
| Slope Trigger |
(>,<, =) Positive slope, (>,<, =) Negative slope
Time: 20ns-10s |
| Alternative Trigger |
CH1 trigger type: Edge, Pulse, Video, Slope
CH2 trigger type: Edge, Pulse, Video, Slope |
| X-Y Mode |
| X-pole Input / Y-Pole Input |
Channel 1 (CH1) / Channel 2 (CH2) |
| Sample Frequency |
XY mode has a breakthrough that trad oscilloscopes restrict sampling rate at 1MSa/s. Support 25Ksa/s~250Msa/s adjusted. |
| Hard Ware Frequency Counter |
| Reading resolution |
1Hz |
| Accuracy |
±0.01% |
| Range |
DC Couple, 10Hz to MAX Bandwidth |
| Signal Types |
Satisfying all Trigger signals(Except Pulse width trigger and Video Trigger) |
| Control Panel Function |
| Auto Set |
Auto adjusting the Vertical, Horizontal system and Trigger Position |
| Save/Recall |
Support 2 Group referenced Waveforms, 20 Group setups, 20 Group captured Waveforms internal Storage/Recall function and USB flash driver storage function. |
| Measure System |
| Auto Measure (32 Types) |
Vpp, Vmax, Vmin, Vamp, Vtop, Vbase, Vavg, Mean,Crms, Vrms, ROVShoot, FOVShoot, RPREShoot, FPREShoot, Rise time, Fall time, Freq, Period,+ Wid,-Wid, +Dut, -Dut, BWid, Phase, FRR, FRF, FFR, FFF, LRR, LRF, LFR, LFF |
| Cursor Measure |
Manual mode, Track mode and Auto mode |