Power Requirement |
AC Voltage |
Selectable 115 or 230V AC RMS |
Frequency |
50/60 Hz |
Mechanical Specification |
Size (HxWxL) |
300mm-200mm-370mm |
Weight |
15kg Approx |
Ground Bond Test – 104, 103, 100 |
Test Voltage |
Nominal 6V AC |
Frequency |
50 or 60 Hz (Independent of supply) |
Display Range and Accuracy |
0-1500mΩ ±2% ±5 Counts |
Display Resolution |
1 mΩ |
Compliance Test Ranges: |
Current/Load Resistance/Time 5 A-1000mΩ - Continuous
10 A-500mΩ - Continuous
25 A-200 mΩ - Continuous
30 A-150 mΩ - 60 secs
40A-100 mΩ - 60 secs
Settable Output Current Range
0.1A-40.0A
Selectable Range of Pass/Fail Levels
0-1500mΩ |
Insulation Resistance Test – 104, 103, 102, 101 |
DC Output Voltage |
0.10kV-6.00kV (10V/Step when manually adjusted)
(50V/Step in test sequence program) |
Display Range |
0.01MΩ-500MΩ |
Display Range / Accuracy |
0.03MΩ-350MΩ ± 5% ± 5 Counts 350MΩ-500MΩ Indication Only |
Display Resolution |
0.01MΩ |
Pass/Fail Level |
0.00MΩ-500.0MΩ |
AC Flash / Hipot Test – 104, 103, 102, 101 |
Programmable Voltage Ranges |
0.10kV-5.00kV (10V/Step when manually adjusted)
(50V/Step in test sequence program) |
Frequency Independent of Supply |
50 or 60 Hz |
Voltage Display Range and Accuracy |
0.10kV-5.00kV ±1% ± 5 Counts |
Voltage Display Resolution |
0.01kV |
Current Display Range and Accuracy |
0.01mA-20.00mA ±1% ± 5 Counts |
Current Display Resolution |
0.01mA |
Selectable Range of Pass/Fail Levels |
0.01mA-20.00mA |
Maximum Current Output |
20.00mA @5kV |
Optional Arc Detection |
9 Levels |
DC Flash / Hipot Test – 104, 103, 102, 101 |
Programmable Voltage Range |
0.10kV-6.00kV (10V/Step when manually adjusted)
(50V/Step in test sequence program) |
Voltage Display Range and Accuracy |
0.10 kV-6.00kV ±1% ± 5 Counts |
Voltage Display Resolution |
0.01kV |
Current Display Range and Accuracy |
0.01mA-10.00mA ±1% ± 5 Counts |
Current Display Resolution |
0.01mA |
Selectable Range of Pass/Fail Levels |
0.01mA-10.00mA |
Maximum Current Output |
10.00mA |
Optional Arc Detection |
9 Levels |
Power Output Rating |
Test Voltage |
110V – 230V AC 20A nominal (Vin-Vout) |
Maximum Power Output |
Up to 5.0kVA (dependant on mains supply rating) |
Output Power Measurement |
Single Phase Power Measurement |
Display Range and Accuracy |
0.02kVA-5.00kVA ±2% ±0.02kVA |
Selectable Range of Pass/Fail Levels |
0.001kVA-5.000kVA |
Single Phase Power Factor Measurement |
Display Range and Accuracy |
Ratio 0.000 -1.000 ± 0.030 |
Selectable Range of Pass/Fail Levels |
0.000-1.000 |
Leakage Current Measurement |
Display Range and Accuracy |
0.10mA-20.00 mA ± 1% ±5 Counts |
Display Resolution |
0.01mA |
Selectable Range of Pass/Fail Levels |
0.01mA-20mA |
Touch Current Measurement |
Display Range and Accuracy |
0.02mA - 5.00 mA ± 1% ±5 Counts |
Display Resolution |
0.01mA |
Selectable Range of Pass/Fail Levels |
0.02mA-5.00mA |
Measuring Device |
IEC/EN 60990 Fig 4
(IEC/EN 60990 Fig 3 or Fig 5 Upon Request) |