General Specifications |
Housing |
ABS/PC/TPE, IP65 |
Dimensions |
297 x 130 x 69 mm |
Weight |
890 g |
Keypad |
Sealed, snap action |
Display |
TFT colour, 480 x 272 pixels, 4.3 inch widescreen, adjustable backlight |
Main processor |
400 MHz ARM |
Memory |
256 MB RAM, 512 MB Flash, SD card 1GB expandable up to 8GB |
Operating system |
Microsoft Windows® CE |
DSP processor |
375 MHz floating point |
Communication |
USB 2.0 |
Power supply |
Rechargeable Lithium-Ion battery pack, 5200 mAh or power adapter |
Battery power |
For min. 16 hours normal use (20°C) |
Operating temperature |
-20 to 50 °C , non condensing |
Charging temperature |
0 to 45 °C |
General features |
Language selection, battery status indication, transducer line test, metric or imperial units |
Meas. point identification |
NFC transponder for communication with CondIDTMtags, read/write distance max. 50 mm (2 inch) |
Vibration monitoring |
Vibration channels |
3 simultaneous |
Dynamic range |
120 dB |
Frequency range |
0 (DC) to 40 kHz |
Resolution |
Max. 25 600 lines |
Vibration transducer input |
<24 Vpp. Transducer supply of 2,5 mA for IEPE(ICP) type can be set On/Off |
Transducer types |
Any transducers (disp., vel. or acc.) with voltage output |
Measuring techniqes |
ISO 2372, ISO 10816, EVAM Evaluated Vibration Analysis, FFT Spectrum with Symptoms, Orbit analysis, 3 channels simultaneously, balancing |
Bearing monitoring |
Measuring range |
SPM HD:-30 to 110 dBsv(44000 transducer)dBm/dBc:-9 to 99 dBsvLR/HR:-19 to 99 dBsv |
Resolution |
0,2 dB/HD, 1 dB dBm/dBc and LR/HR |
Transducer types |
SPM 40000, 42000, 44000, probe and quick connector transducers |
Tacho input |
Measuring range |
1 to 150 000 PPM |
Resolution |
1 pulse |
Accuracy |
± (1 pulse + 0.01% of reading) |
Transducer types |
TTP10, TTL pulses, keyphasor and proximity switch NPN/PNP. |
Output |
TTL output for stroboscope and 12 VDC |
Analog signals |
Measurement range |
0 to 10 V DC, 0 to 20 mA |
Resolution |
18 bit A/D converter |
Accuracy |
± 1% of reading + 0,1 V/mA |
Output/input |
Headphones/microphone |
3.5 mm stereo plug |
Communication |
Mini USB |
Temperature measurement |
Inputs |
TTP10 or via analog input |
Stethoscope |
Transducer types |
Shock pulse and vibration transducer |
Settings |
Filter, volume and gain |