| INPUT |
| Voltage |
100 - 120Vac / 200 - 240Vac±10% Auto Range |
| Frequency |
50/60Hz ± 5% |
| Fuse |
15A / 250Vac Fast-Blow |
| AC WITHSTAND VOLTAGE |
| |
Range |
Resolution |
Accuracy |
| Output Voltage, KVAC |
0-5.00 |
0.01 |
± (2% of setting + 5V) |
| Output Frequency |
50Hz/60Hz ± 0.1% , User Selection |
| Output Waveform |
Sine Wave, Crest Factor = 1.3 - 1.5 and output voltage > 300V |
| Output Regulation, Vrms |
± (1% of output + 5V), From no load to full load and Low Line to High Line (combined regulation) |
| Hi-Limit AC current, mA |
0.10-99.99 |
0.01 |
± (2% of setting + 6 counts) |
| Lo-Limit AC current, mA |
0.00-99.99 |
0.01 |
± (2% of setting + 6 counts) |
| Ramp Time, second |
0.2-180.0 |
0.1 |
± (0.1% of setting + 0.05 sec) |
| Dwell Time, second |
0, 0.2 - 60.0 (0=continuous) |
0.1 |
± (0.1% of setting + 0.05 sec) |
| DC WITHSTAND VOLTAGE |
| |
Range |
Resolution |
Accuracy |
| Output Voltage, KVDC |
0 -6.00 |
0.01 |
± (2% of setting + 5V) |
| DC Output Ripple, Vrms |
<5 % (6KV / 10mA at Resistive Load) |
| Hi-Limit DC current, mA |
0.02 - 10.00 |
0.01 |
± (2% of setting + 2 counts) |
| Lo-Limit DC current, mA |
0.00 - 10.00 |
0.01 |
± (2% of setting + 2 counts) |
| Ramp Time, second |
0.2 - 180.0 |
0.1 |
± (0.1% of setting + 0.05 sec) |
| Dwell Time, second |
0, 0.2 - 60.0 (0=continuous) |
0.1 |
± (0.1% of setting + 0.05 sec) |
| Discharge Time |
< 50 msec for no load, < 100 msec for capacitor load |
| Maximum Capacitive Load DC Mode |
1µF < 1KV, 0.08µF < 4KV
0.75µF < 2KV, 0.04µF < 5KV
0.5µF < 3KV, 0.015uF < 6KV |
| INSULATION RESISTANCE |
| |
Range |
Resolution |
Accuracy |
| Output Voltage, VDC |
100 - 1000 |
1 |
± (2% of setting + 5V) |
| Hi-Limit resistance, MΩ |
0, 1 - 1000 0 = OFF) |
1 |
100-499V ± (7% of setting + 2 counts) |
| Lo-Limit resistance, MΩ |
1 - 1000 |
1 |
500-1000V ± (3% of setting + 2 counts) |
| Ramp Time, second |
0.1 or 2.0 |
0.1 |
± (0.1% of setting + 0.05 sec) |
| Delay Time, second |
0, 0.5 - 999.9 |
0.1 |
± (0.1% of setting + 0.05 sec) |
| GROUND BOND |
| |
Range |
Resolution |
Accuracy |
| Output AC Current, A |
1.0 - 40.0 |
0.1 |
± (2 % of setting + 2 counts) |
| Output AC Voltage, V |
8V(Fixed) |
| Output Frequency, Hz |
50Hz/60Hz ± 0.1%, User Selectable |
| Maximum Loading |
1.0-10.0A/0-600mΩ,10.1-30.0A/0-200mΩ,30.1-40.0A/0-150mΩ |
| Offset, mΩ |
0-100 |
1 |
± (2 % of setting + 2 counts) |
| HI and LO-Limit Resistance, mΩ |
0 - 150 (30.1-40.0A) |
1 |
± (2 % of setting + 2 counts) |
| 0 - 200 (10.1-30.0A) |
| 0 - 600 (1.0-10.0A) |
| Fix Ramp Timer, second |
0.4 |
0.1 |
± 0.05 sec |
0-250mOhm à 0.1sec ramp up
251-300mOhm à 0.2sec ramp up
301-450mOhm à 0.3sec ramp up
>450mOhm = 0.4sec |
| Dwell Timer, second |
0, 0.1 - 240.0 (0 = continuous) |
0.1 |
± 0.05 sec |
| GENERAL SPECIFICATIONS |
| Memories |
Allows storage of up to 20 different test programs and a single step mode |
| Remote I/O |
Input: Test, Reset, Interlock & Recall Memory 1 - 6
Output: Pass, Fail, Test-in-Process |
| Interface |
USB (Optional) |
| Security |
Lockout capability to avoid unauthorized access to test set-up programs |
| Calibration |
Software and adjustments made through the front panel |
| Dimension |
430mm(W) ×132mm(H) × 400 mm(D) |
| Weight |
53lbs (24kg) |