Range |
with R&S®SMCVB-B103 option (mandatory) |
4 kHz to 3 GHz |
Resolution of setting |
|
0.001 Hz |
Resolution of synthesis |
f = 1 GHz |
2.665 μHz (nom.) |
Settling time |
to within < 1 × 10-7 for f > 200 MHz
or < 20 Hz for f ≤ 200 MHz
with GUI update stopped, I/Q optimization
mode: fast, measured from command at
instrument to frequency settled within
specified range, with Ethernet (fast socket) remote control,
level setting characteristic: auto |
< 5 ms |
Range and resolution of phase offset setting |
|
-999.99° to +999.99°, 0.01° resolution |
Reference frequency |
Frequency error |
at time of calibration in production |
< 1 × 10-7 |
Aging |
after 30 days of uninterrupted operation |
≤ 1 × 10-6 /year |
Temperature effect |
in temperature range from +5 °C to +45 °C |
±1.0 × 10-6 |
Source |
|
internal, external |
External reference frequency modes |
standard |
10 MHz |
Reference frequency input |
Connector type |
REF IN on rear panel |
BNC female |
Input frequency |
|
10 MHz, 13 MHz |
Minimum frequency locking range |
|
± 25 × 10-6 (meas.) |
Input level range |
|
0 dBm to +16 dBm (meas.) |
Input impedance |
|
50 Ω (nom.) |
Reference frequency output |
Connector type |
REF OUT on rear panel |
BNC female |
Output frequency |
square wave |
|
source mode: internal |
10 MHz |
|
source mode: external |
10 MHz |
Output level |
|
+7 dBm to +13 dBm, +10 dBm (meas.) |
Source impedance |
|
50 Ω (nom.) |
Setting range |
R&S®SMCVB-B103/-KB106/-KB107 |
standard |
|
4 kHz ≤ f < 100 kHz |
-120 dBm to +16 dBm |
|
100 kHz ≤ f < 6 GHz |
-145 dBm to +16 dBm |
|
6 GHz ≤ f ≤ 7.125 GHz |
-145 dBm to +16 dBm |
|
with R&S®SMCVB-K31 option |
|
4 kHz ≤ f < 100 kHz |
-120 dBm to +25 dBm |
|
100 kHz ≤ f ≤ 6 GHz |
-145 dBm to +25 dBm |
|
6 GHz ≤ f ≤ 7.125 GHz |
-145 dBm to +25 dBm |
Setting resolution |
|
0.01 dB |
Specified level range |
peak envelope power (PEP) |
R&S®SMCVB-B103/-KB106/-KB107 |
standard |
|
4 kHz < f ≤ 10 MHz |
-110 dBm to +15 dBm |
|
10 MHz < f ≤ 6 GHz 1 |
-120 dBm to +15 dBm |
|
6 GHz < f ≤ 7.125 GHz |
-120 dBm to +15 dBm |
|
with R&S®SMCVB-K31 option |
|
4 kHz < f ≤ 10 MHz |
-110 dBm to +20 dBm |
|
10 MHz < f ≤ 6 GHz1 |
-120 dBm to +20 dBm |
|
6 GHz < f ≤ 7.125 GHz |
-120 dBm to +18 dBm |
Level accuracy |
level setting characteristic: auto, temperature range from +18 °C to +33 °C |
|
level > -80 dBm |
|
4 kHz < f < 200 kHz |
< 1.8 dB |
|
200 kHz ≤ f ≤ 10 MHz |
< 0.7 dB |
|
10 MHz < f ≤ 2.5 GHz 1 |
< 0.5 dB |
|
f > 2.5 GHz 1 |
< 0.7 dB |
|
level ≤ -80 dBm |
|
4 kHz < f < 200 kHz |
< 1.8 dB |
|
200 kHz ≤ f ≤ 10 MHz |
< 1.2 dB, < 1.0 dB (typ.) |
|
10 MHz < f ≤ 2.5 GHz 1 |
< 0.8 dB |
|
f > 2.5 GHz 1 |
< 1.1 dB |
Settling time |
to < 0.1 dB deviation from final value,
with GUI update stopped,
temperature range from +18 °C to +33 °C,
f > 10 MHz, I/Q optimization mode: fast,
measured from command at instrument to
frequency settled within specified range,
with Ethernet (fast socket) remote control,
level setting characteristic: auto |
< 5 ms |
Interruption-free level range |
level setting characteristic: uninterrupted level setting |
> 20 dB |